Reference intervals for reticulocyte count and derived reticulocyte parameters in a cohort of healthy adults
Introduction Reticulocyte count and novel derived parameters provide insight into the effectiveness of erythropoiesis and may be useful tools in the classification and diagnosis of anemias. However, there is no standardisation, so we consider it necessary that each laboratory evaluates the parameter...
Saved in:
Published in: | International journal of laboratory hematology 2024-12, Vol.46 (6), p.997-1003 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Introduction
Reticulocyte count and novel derived parameters provide insight into the effectiveness of erythropoiesis and may be useful tools in the classification and diagnosis of anemias. However, there is no standardisation, so we consider it necessary that each laboratory evaluates the parameters according to its own methodology and instrumentation and establishes its own reference ranges. Our aim was to establish the reference intervals (RIs) of reticulocyte profile provided by the Beckman Coulter DxH 900 haematological autoanalyzer in our reference population.
Methods
One hundred and seventy‐five healthy adults (18 to 62 years) were included. Subjects were collected from the blood donation centre of the Hospital Clínico San Carlos (Madrid, Spain) upon informed consent. Whole blood was collected and assayed for 14 haematological parameters on the Beckman Coulter DxH 900 analyzer in the haematology laboratory of the Clinical Analysis Department. RIs were established as per Clinical and Laboratory Standards Institute EP28‐A3c guidelines using three different statistical approaches.
Results
RIs estimated using the non‐parametric method and the Harrell‐Davis bootstrap method were very similar. RIs estimated by the robust method were narrower. Gender partitioning was required for two haematological parameters (low haemoglobin density (LHD) and microcytic anaemia factor (MAF)). The rest of the parameters did not need to be partitioned according to Lahti's method.
Conclusion
RIs have been established for 14 hematologic parameters of the reticulocyte profile for the Beckman Coulter DxH 900 haematology analyzer using a healthy cohort of adult subjects. |
---|---|