Loading…

Insertable, dual-density dielectric barrier for acoustic pressure level reduction in a high-performance human head-only MRI system

We report use of a dual-density dielectric barrier surrounding a detachable high-pass radiofrequency (RF) birdcage coil to achieve an order-of-magnitude reduction of acoustic noise in a high-performance head gradient system. The barrier consisted of a 4.5 mm-thick mass-loaded vinyl and a 6 mm-thick...

Full description

Saved in:
Bibliographic Details
Published in:Magnetic resonance imaging 2024-12, Vol.114, p.110224, Article 110224
Main Authors: Lee, S.-K., Tarasek, Matthew R., Park, Keith, Yeo, Desmond T.-B., Foo, Thomas K.-F.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report use of a dual-density dielectric barrier surrounding a detachable high-pass radiofrequency (RF) birdcage coil to achieve an order-of-magnitude reduction of acoustic noise in a high-performance head gradient system. The barrier consisted of a 4.5 mm-thick mass-loaded vinyl and a 6 mm-thick polyurethane foam. It was inserted into the radial gap between the birdcage coil and the RF shield in a prototype head-only gradient system at 3 T. More than 9 dBA reduction of sound pressure level was achieved on the average with representative, high acoustic-noise imaging sequences. Increased acoustic damping was apparent from acoustic impulse response functions. High dielectric constant of the mass-loaded vinyl effectively added distributed capacitance to the birdcage coil, lowering the resonance frequency, but not seriously degrading the RF transmission performance. The barrier occupied the radial space normally used for air cooling of the RF coil and the RF shield. The resulting omission of air cooling was found to be acceptable with efficient gradient thermal management and use of a high-resistivity RF shield for eddy current reduction. The proposed method can improve patient experience while preserving image quality in a high-power head-only gradient system.
ISSN:0730-725X
1873-5894
1873-5894
DOI:10.1016/j.mri.2024.110224