Loading…

Comparison of transepidermal water loss (TEWL) measurements with two novel sensors based on different sensing principles

In this paper we compare results of TEWL measurements carried out with two different instruments recently developed in our institutes. One sensor is based on a dew point hygrometer, while the other one uses a conductance type humidity sensor. In both cases the closed chamber measuring method is used...

Full description

Saved in:
Bibliographic Details
Published in:Sensors and actuators. A. Physical. 2008-03, Vol.142 (1), p.67-72
Main Authors: Mündlein, Martin, Valentin, Bernhard, Chabicovsky, Rupert, Nicolics, Johann, Weremczuk, Jerzy, Tarapata, Grzegorz, Jachowicz, Ryszard
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In this paper we compare results of TEWL measurements carried out with two different instruments recently developed in our institutes. One sensor is based on a dew point hygrometer, while the other one uses a conductance type humidity sensor. In both cases the closed chamber measuring method is used. Furthermore, we have applied the same principle of calibration to both instruments.
ISSN:0924-4247
1873-3069
DOI:10.1016/j.sna.2007.04.012