Measured and Calculated Doses of Transmitted and Reflected Radiation in the Energy Range of X-Ray Diagnostic Apparatuses
Measured and calculated doses of transmitted and reflected X ray radiation were compared. Calculated doses were obtained using the Monte Carlo program GEANT and a model of X ray radiation transmission/reflection. X Ray radiation load on the patient was decreased using filtering of soft (non-working)...
Saved in:
Published in: | Biomedical engineering 2008-09, Vol.42 (5), p.243-246 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Measured and calculated doses of transmitted and reflected X ray radiation were compared. Calculated doses were obtained using the Monte Carlo program GEANT and a model of X ray radiation transmission/reflection. X Ray radiation load on the patient was decreased using filtering of soft (non-working) range of the X ray spectrum ( |
---|---|