Loading…
Pilot Quality Control Program for Audit RT External Beams at Mexican Hospitals
A pilot quality control program for audit 18 radiotherapy RT external beams at 13 Mexican hospitals is described—for eleven 60 Co beams and seven photon beams of 6, 10 and 15 MV from accelerators. This program contains five parts: a) Preparation of the TLD‐100 powder: washing, drying and annealing (...
Saved in:
Main Authors: | , |
---|---|
Format: | Conference Proceeding |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A pilot quality
control program for audit 18 radiotherapy RT external beams at 13 Mexican hospitals is
described—for eleven
60
Co
beams and seven photon beams of 6, 10 and 15 MV from accelerators. This
program contains five parts: a) Preparation of the TLD‐100 powder: washing, drying and
annealing (one
hour 400 °C plus 24 hrs 80 °C). b) Sending two IAEA type capsules to the hospitals for
irradiation at the hospital to a nominal
D
W
= 2
Gy
⋅
c
) Preparation at the SSDL of ten calibration curves CC in the
range of 0.5 Gy to 6 Gy in terms of absorbed dose to water
D
W
for
60
Co
with traceability to primary laboratory NRC (Canada),
according to a window irradiation: 26/10/2007–7/12/2007. d) Reading all capsules that
match their hospital time irradiation and the SSDL window irradiation. f) Evaluation of
the Dw imparted by the hospitals. |
---|---|
ISSN: | 0094-243X 1551-7616 |
DOI: | 10.1063/1.2979268 |