Loading…

First application of Cc-corrected imaging for high-resolution and energy-filtered TEM

Contrast-transfer calculations indicate that Cc correction should be highly beneficial for high-resolution and energy-filtered transmission electron microscopy. A prototype of an electron optical system capable of correcting spherical and chromatic aberration has been used to verify these calculatio...

Full description

Saved in:
Bibliographic Details
Published in:Journal of electron microscopy 2009-06, Vol.58 (3), p.147-155
Main Authors: Kabius, Bernd, Hartel, Peter, Haider, Maximilian, Müller, Heiko, Uhlemann, Stephan, Loebau, Ulrich, Zach, Joachim, Rose, Harald
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Contrast-transfer calculations indicate that Cc correction should be highly beneficial for high-resolution and energy-filtered transmission electron microscopy. A prototype of an electron optical system capable of correcting spherical and chromatic aberration has been used to verify these calculations. A strong improvement in resolution at an acceleration voltage of 80 kV has been measured. Our first Cc-corrected energy-filtered experiments examining a (LaAlO3)0.3(Sr2AlTaO6)0.7/LaCoO3 interface demonstrated a significant gain for the spatial resolution in elemental maps of La.