Efficacy of distinct energy delivery protocols comparing two biphasic defibrillators for cardiac arrest

Limited data have been published on the use of external defibrillators that deliver impedance compensated biphasic (ICB) waveforms in patients. We compared 2 ICB defibrillators, the Heartstream XL (150-150-150 J protocol) and Heartsine Samaritan (100-150-200 J protocol) in 78 consecutive patients in...

Full description

Saved in:
Bibliographic Details
Published in:The American journal of cardiology 2004-08, Vol.94 (3), p.378-380
Main Authors: Walsh, Simon J, McClelland, Anthony J.J, Owens, Colum G, Allen, James, Anderson, John McC, Turner, Colin, Adgey, A.A.Jennifer
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Limited data have been published on the use of external defibrillators that deliver impedance compensated biphasic (ICB) waveforms in patients. We compared 2 ICB defibrillators, the Heartstream XL (150-150-150 J protocol) and Heartsine Samaritan (100-150-200 J protocol) in 78 consecutive patients in cardiac arrest. The performance of the 2 devices over the first 2 shocks was statistically equivalent. By the third shock, the Heartsine Samaritan had significantly better performance in removing ventricular fibrillation (p = 0.029). Energy selection for ICB waveforms requires further validation.