Fracture of a Cross-Linked Polyethylene Liner Due to Impingement

Abstract We report a case of fracture at 2 years after implantation of a 50-kGy moderately cross-linked ultrahigh molecular weight polyethylene liner with an extended lip (Marathon, DePuy, Warsaw, IN). The extended lip section had fractured. The liner showed no oxidation. The articular surface was g...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of arthroplasty 2009, Vol.24 (1), p.158.e15-158.e19
Main Authors: Duffy, Gavan P., MD, Wannomae, Keith K., BS, Rowell, Shannon L., BS, Muratoglu, Orhun K., PhD
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3
cites cdi_FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3
container_end_page 158.e19
container_issue 1
container_start_page 158.e15
container_title The Journal of arthroplasty
container_volume 24
creator Duffy, Gavan P., MD
Wannomae, Keith K., BS
Rowell, Shannon L., BS
Muratoglu, Orhun K., PhD
description Abstract We report a case of fracture at 2 years after implantation of a 50-kGy moderately cross-linked ultrahigh molecular weight polyethylene liner with an extended lip (Marathon, DePuy, Warsaw, IN). The extended lip section had fractured. The liner showed no oxidation. The articular surface was grossly deformed, likely due to wear, creep, and/or plastic deformation, and the liner showed no recovery of machining marks upon melting, indicating that some wear had occurred. Electron microscopy revealed fatigue striations on the fracture surface. The likely cause of failure was femoral neck impingement–induced wear and fatigue on the liner.
doi_str_mv 10.1016/j.arth.2007.12.020
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_66751957</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>1_s2_0_S0883540308000089</els_id><sourcerecordid>66751957</sourcerecordid><originalsourceid>FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3</originalsourceid><addsrcrecordid>eNp9kU2LFDEQhoMo7rj6BzxIn7x1W0m68wEiLqOrCwO7sHoO6aTazWx_jEn3wvx708yA4MFcCsLzvlBPEfKWQkWBig_7ysb5oWIAsqKsAgbPyIY2nJWqBvGcbEApXjY18AvyKqU9AKVNU78kF1RTAM3Ehny-jtbNS8Ri6gpbbOOUUrkL4yP64m7qjzg_HHscsch_GIsvCxbzVNwMhzD-wgHH-TV50dk-4ZvzvCQ_r7_-2H4vd7ffbrZXu9JxDXMpWmlbXkMt0QutAYRUKIV2TklBoc1QLbxGpa3vtNSUdY3iumFSc-l9yy_J-1PvIU6_F0yzGUJy2Pd2xGlJRgjZUN3IDLIT6NZdInbmEMNg49FQMKs3szerN7N6M5SZ7C2H3p3bl3ZA_zdyFpWBjycA845PAaNJLuDo0IeIbjZ-Cv_v__RP3PVhDM72j3jEtJ-WOGZ7hpqUA-Z-vdx6OFCQn9L8DwU4kXA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>66751957</pqid></control><display><type>article</type><title>Fracture of a Cross-Linked Polyethylene Liner Due to Impingement</title><source>ScienceDirect Freedom Collection</source><creator>Duffy, Gavan P., MD ; Wannomae, Keith K., BS ; Rowell, Shannon L., BS ; Muratoglu, Orhun K., PhD</creator><creatorcontrib>Duffy, Gavan P., MD ; Wannomae, Keith K., BS ; Rowell, Shannon L., BS ; Muratoglu, Orhun K., PhD</creatorcontrib><description>Abstract We report a case of fracture at 2 years after implantation of a 50-kGy moderately cross-linked ultrahigh molecular weight polyethylene liner with an extended lip (Marathon, DePuy, Warsaw, IN). The extended lip section had fractured. The liner showed no oxidation. The articular surface was grossly deformed, likely due to wear, creep, and/or plastic deformation, and the liner showed no recovery of machining marks upon melting, indicating that some wear had occurred. Electron microscopy revealed fatigue striations on the fracture surface. The likely cause of failure was femoral neck impingement–induced wear and fatigue on the liner.</description><identifier>ISSN: 0883-5403</identifier><identifier>EISSN: 1532-8406</identifier><identifier>DOI: 10.1016/j.arth.2007.12.020</identifier><identifier>PMID: 19100926</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Aged, 80 and over ; Arthralgia - diagnosis ; Arthralgia - etiology ; Arthroplasty, Replacement, Hip - adverse effects ; Arthroplasty, Replacement, Hip - instrumentation ; early failure ; fatigue failure ; Female ; Femur Neck ; highly cross-linked polyethylene ; Hip Joint - diagnostic imaging ; Hip Joint - surgery ; Hip Prosthesis ; Humans ; impingement ; Orthopedics ; Polyethylene ; Prosthesis Failure ; Radiography ; total hip arthroplasty</subject><ispartof>The Journal of arthroplasty, 2009, Vol.24 (1), p.158.e15-158.e19</ispartof><rights>Elsevier Inc.</rights><rights>2009 Elsevier Inc.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3</citedby><cites>FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,4010,27897,27898,27899</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/19100926$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Duffy, Gavan P., MD</creatorcontrib><creatorcontrib>Wannomae, Keith K., BS</creatorcontrib><creatorcontrib>Rowell, Shannon L., BS</creatorcontrib><creatorcontrib>Muratoglu, Orhun K., PhD</creatorcontrib><title>Fracture of a Cross-Linked Polyethylene Liner Due to Impingement</title><title>The Journal of arthroplasty</title><addtitle>J Arthroplasty</addtitle><description>Abstract We report a case of fracture at 2 years after implantation of a 50-kGy moderately cross-linked ultrahigh molecular weight polyethylene liner with an extended lip (Marathon, DePuy, Warsaw, IN). The extended lip section had fractured. The liner showed no oxidation. The articular surface was grossly deformed, likely due to wear, creep, and/or plastic deformation, and the liner showed no recovery of machining marks upon melting, indicating that some wear had occurred. Electron microscopy revealed fatigue striations on the fracture surface. The likely cause of failure was femoral neck impingement–induced wear and fatigue on the liner.</description><subject>Aged, 80 and over</subject><subject>Arthralgia - diagnosis</subject><subject>Arthralgia - etiology</subject><subject>Arthroplasty, Replacement, Hip - adverse effects</subject><subject>Arthroplasty, Replacement, Hip - instrumentation</subject><subject>early failure</subject><subject>fatigue failure</subject><subject>Female</subject><subject>Femur Neck</subject><subject>highly cross-linked polyethylene</subject><subject>Hip Joint - diagnostic imaging</subject><subject>Hip Joint - surgery</subject><subject>Hip Prosthesis</subject><subject>Humans</subject><subject>impingement</subject><subject>Orthopedics</subject><subject>Polyethylene</subject><subject>Prosthesis Failure</subject><subject>Radiography</subject><subject>total hip arthroplasty</subject><issn>0883-5403</issn><issn>1532-8406</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2009</creationdate><recordtype>article</recordtype><recordid>eNp9kU2LFDEQhoMo7rj6BzxIn7x1W0m68wEiLqOrCwO7sHoO6aTazWx_jEn3wvx708yA4MFcCsLzvlBPEfKWQkWBig_7ysb5oWIAsqKsAgbPyIY2nJWqBvGcbEApXjY18AvyKqU9AKVNU78kF1RTAM3Ehny-jtbNS8Ri6gpbbOOUUrkL4yP64m7qjzg_HHscsch_GIsvCxbzVNwMhzD-wgHH-TV50dk-4ZvzvCQ_r7_-2H4vd7ffbrZXu9JxDXMpWmlbXkMt0QutAYRUKIV2TklBoc1QLbxGpa3vtNSUdY3iumFSc-l9yy_J-1PvIU6_F0yzGUJy2Pd2xGlJRgjZUN3IDLIT6NZdInbmEMNg49FQMKs3szerN7N6M5SZ7C2H3p3bl3ZA_zdyFpWBjycA845PAaNJLuDo0IeIbjZ-Cv_v__RP3PVhDM72j3jEtJ-WOGZ7hpqUA-Z-vdx6OFCQn9L8DwU4kXA</recordid><startdate>2009</startdate><enddate>2009</enddate><creator>Duffy, Gavan P., MD</creator><creator>Wannomae, Keith K., BS</creator><creator>Rowell, Shannon L., BS</creator><creator>Muratoglu, Orhun K., PhD</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>2009</creationdate><title>Fracture of a Cross-Linked Polyethylene Liner Due to Impingement</title><author>Duffy, Gavan P., MD ; Wannomae, Keith K., BS ; Rowell, Shannon L., BS ; Muratoglu, Orhun K., PhD</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2009</creationdate><topic>Aged, 80 and over</topic><topic>Arthralgia - diagnosis</topic><topic>Arthralgia - etiology</topic><topic>Arthroplasty, Replacement, Hip - adverse effects</topic><topic>Arthroplasty, Replacement, Hip - instrumentation</topic><topic>early failure</topic><topic>fatigue failure</topic><topic>Female</topic><topic>Femur Neck</topic><topic>highly cross-linked polyethylene</topic><topic>Hip Joint - diagnostic imaging</topic><topic>Hip Joint - surgery</topic><topic>Hip Prosthesis</topic><topic>Humans</topic><topic>impingement</topic><topic>Orthopedics</topic><topic>Polyethylene</topic><topic>Prosthesis Failure</topic><topic>Radiography</topic><topic>total hip arthroplasty</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Duffy, Gavan P., MD</creatorcontrib><creatorcontrib>Wannomae, Keith K., BS</creatorcontrib><creatorcontrib>Rowell, Shannon L., BS</creatorcontrib><creatorcontrib>Muratoglu, Orhun K., PhD</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of arthroplasty</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Duffy, Gavan P., MD</au><au>Wannomae, Keith K., BS</au><au>Rowell, Shannon L., BS</au><au>Muratoglu, Orhun K., PhD</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Fracture of a Cross-Linked Polyethylene Liner Due to Impingement</atitle><jtitle>The Journal of arthroplasty</jtitle><addtitle>J Arthroplasty</addtitle><date>2009</date><risdate>2009</risdate><volume>24</volume><issue>1</issue><spage>158.e15</spage><epage>158.e19</epage><pages>158.e15-158.e19</pages><issn>0883-5403</issn><eissn>1532-8406</eissn><abstract>Abstract We report a case of fracture at 2 years after implantation of a 50-kGy moderately cross-linked ultrahigh molecular weight polyethylene liner with an extended lip (Marathon, DePuy, Warsaw, IN). The extended lip section had fractured. The liner showed no oxidation. The articular surface was grossly deformed, likely due to wear, creep, and/or plastic deformation, and the liner showed no recovery of machining marks upon melting, indicating that some wear had occurred. Electron microscopy revealed fatigue striations on the fracture surface. The likely cause of failure was femoral neck impingement–induced wear and fatigue on the liner.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>19100926</pmid><doi>10.1016/j.arth.2007.12.020</doi></addata></record>
fulltext fulltext
identifier ISSN: 0883-5403
ispartof The Journal of arthroplasty, 2009, Vol.24 (1), p.158.e15-158.e19
issn 0883-5403
1532-8406
language eng
recordid cdi_proquest_miscellaneous_66751957
source ScienceDirect Freedom Collection
subjects Aged, 80 and over
Arthralgia - diagnosis
Arthralgia - etiology
Arthroplasty, Replacement, Hip - adverse effects
Arthroplasty, Replacement, Hip - instrumentation
early failure
fatigue failure
Female
Femur Neck
highly cross-linked polyethylene
Hip Joint - diagnostic imaging
Hip Joint - surgery
Hip Prosthesis
Humans
impingement
Orthopedics
Polyethylene
Prosthesis Failure
Radiography
total hip arthroplasty
title Fracture of a Cross-Linked Polyethylene Liner Due to Impingement
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-25T16%3A00%3A37IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Fracture%20of%20a%20Cross-Linked%20Polyethylene%20Liner%20Due%20to%20Impingement&rft.jtitle=The%20Journal%20of%20arthroplasty&rft.au=Duffy,%20Gavan%20P.,%20MD&rft.date=2009&rft.volume=24&rft.issue=1&rft.spage=158.e15&rft.epage=158.e19&rft.pages=158.e15-158.e19&rft.issn=0883-5403&rft.eissn=1532-8406&rft_id=info:doi/10.1016/j.arth.2007.12.020&rft_dat=%3Cproquest_cross%3E66751957%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c390t-6b7ab34047ed69900678e769cc87610b39046d9e89adf97912f5839527937ddb3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=66751957&rft_id=info:pmid/19100926&rfr_iscdi=true