Loading…
Chiral separation of β-adrenergic blockers on CelluCoat column by HPLC
The chiral separation of 10 β-adrenergic blockers (acebutalol, alprenolol, bufuralol, bisoprolol, celiprolol, carazolol, indenolol, metoprolol, oxprenolol and propranolol) was achieved on CelluCoat column (250 mm × 4.6 mm, 5 μm particle size). The mobile phases used were (90:10:0.2, v/v/v) and (95:5...
Saved in:
Published in: | Talanta (Oxford) 2009-04, Vol.78 (2), p.458-463 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The chiral separation of 10 β-adrenergic blockers (acebutalol, alprenolol, bufuralol, bisoprolol, celiprolol, carazolol, indenolol, metoprolol, oxprenolol and propranolol) was achieved on CelluCoat column (250
mm
×
4.6
mm, 5
μm particle size). The mobile phases used were (90:10:0.2, v/v/v) and (95:5:0.2, v/v/v) combinations of
n-heptane–ethanol–diethylamine, respectively. The flow rates were 0.5, 1.0 and 2.0
mL
min
−1 with detection at 225
nm. The capacity (
k), selectivity (
α) and resolution (
R
s) factors were 0.44–12.91, 1.12–2.19 and 1.00–9.50, respectively. The proposed supra-molecular models indicated that the chiral resolution were governed by π–π interactions, hydrogen bondings and steric effect. |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/j.talanta.2008.11.043 |