Loading…

Acute Liner Disassociation of a Pinnacle Acetabular Component

Abstract A primary total hip arthroplasty using a cementless Pinnacle cup and neutral Marathon polyethylene liner mated via a 32-mm cobalt chrome head to a Summit pore coated stem (Depuy, Warsaw, IN). At 23 months postoperative an acute disassociation occurred. This was resolved with a liner exchang...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of arthroplasty 2009-08, Vol.24 (5), p.815-818
Main Author: Mesko, J. Wesley, MD
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract A primary total hip arthroplasty using a cementless Pinnacle cup and neutral Marathon polyethylene liner mated via a 32-mm cobalt chrome head to a Summit pore coated stem (Depuy, Warsaw, IN). At 23 months postoperative an acute disassociation occurred. This was resolved with a liner exchange to a 36-mm metal liner with a 36-mm metal head. The retrieved polyethylene liner was noted to have 3 of the 6 peripheral locking tabs sheared off.