Loading…
Cadaver validation of intensity-based ultrasound to CT registration
A method is presented for the rigid registration of tracked B-mode ultrasound images to a CT volume of a femur and pelvis. This registration can allow tracked surgical instruments to be aligned with the CT image or an associated preoperative plan. Our method is fully automatic and requires no manual...
Saved in:
Published in: | Medical image analysis 2006-06, Vol.10 (3), p.385-395 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A method is presented for the rigid registration of tracked B-mode ultrasound images to a CT volume of a femur and pelvis. This registration can allow tracked surgical instruments to be aligned with the CT image or an associated preoperative plan. Our method is fully automatic and requires no manual segmentation of either the ultrasound images or the CT volume. The parameter which is directly related to the speed of sound through tissue has also been included in the registration optimisation process. Experiments have been carried out on six cadaveric femurs and three cadaveric pelves. Registration results were compared with a “gold standard” registration acquired using bone implanted fiducial markers. Results show the registration method to be accurate, on average, to 1.6
mm root-mean-square target registration error. |
---|---|
ISSN: | 1361-8415 1361-8423 |
DOI: | 10.1016/j.media.2006.01.003 |