Loading…

Cadaver validation of intensity-based ultrasound to CT registration

A method is presented for the rigid registration of tracked B-mode ultrasound images to a CT volume of a femur and pelvis. This registration can allow tracked surgical instruments to be aligned with the CT image or an associated preoperative plan. Our method is fully automatic and requires no manual...

Full description

Saved in:
Bibliographic Details
Published in:Medical image analysis 2006-06, Vol.10 (3), p.385-395
Main Authors: Penney, G.P., Barratt, D.C., Chan, C.S.K., Slomczykowski, M., Carter, T.J., Edwards, P.J., Hawkes, D.J.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A method is presented for the rigid registration of tracked B-mode ultrasound images to a CT volume of a femur and pelvis. This registration can allow tracked surgical instruments to be aligned with the CT image or an associated preoperative plan. Our method is fully automatic and requires no manual segmentation of either the ultrasound images or the CT volume. The parameter which is directly related to the speed of sound through tissue has also been included in the registration optimisation process. Experiments have been carried out on six cadaveric femurs and three cadaveric pelves. Registration results were compared with a “gold standard” registration acquired using bone implanted fiducial markers. Results show the registration method to be accurate, on average, to 1.6 mm root-mean-square target registration error.
ISSN:1361-8415
1361-8423
DOI:10.1016/j.media.2006.01.003