Loading…

Method of calibration to correct for cloud-induced wavelength shifts in the Aura satellite's Ozone Monitoring Instrument

The in-flight wavelength calibration for the Ozone Monitoring Instrument is discussed. The observed variability in the wavelength scale is two orders of magnitude larger than caused by temperature changes in the instrument. These wavelength variations are the result of rapid changes in time in the r...

Full description

Saved in:
Bibliographic Details
Published in:Applied optics (2004) 2006-05, Vol.45 (15), p.3652-3658
Main Authors: Voors, Robert, Dobber, Marcel, Dirksen, Ruud, Levelt, Pieternel
Format: Article
Language:English
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The in-flight wavelength calibration for the Ozone Monitoring Instrument is discussed. The observed variability in the wavelength scale is two orders of magnitude larger than caused by temperature changes in the instrument. These wavelength variations are the result of rapid changes in time in the radiance levels during an individual observation in the presence of clouds or snow and ice. We have developed a data processing method to account and correct for these changes. In February 2005 this correction was implemented in the official data processing stream. We explain in detail how and how accurately this method works. Before correction, the error in the wavelength scale can be as much as a few tenths of a pixel; after correction it is mostly less than 1/100th of a pixel, which is the required preflight accuracy. This means that higher-level products such as the total column amounts of ozone, NO2, and SO2 are not significantly affected. It is expected that these wavelength variations will be observed in other hyperspectral Earth observation spectrometers and that the correction mechanism should apply equally well.
ISSN:1559-128X
DOI:10.1364/ao.45.003652