Loading…
Potential benefit of transesophageal defibrillation: an experimental evaluation
Because of the proximity of the esophagus to the heart, transesophageal defibrillation might increase defibrillation success. We assessed the defibrillation threshold (DFT) of transesophageal defibrillation compared with standard transthoracic defibrillation. Defibrillation success and DFTs were det...
Saved in:
Published in: | The American journal of emergency medicine 2006-07, Vol.24 (4), p.418-422 |
---|---|
Main Authors: | , , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Because of the proximity of the esophagus to the heart, transesophageal defibrillation might increase defibrillation success. We assessed the defibrillation threshold (DFT) of transesophageal defibrillation compared with standard transthoracic defibrillation.
Defibrillation success and DFTs were determined in 22 female pigs with high (68 ± 4 kg, n = 12) or low body weight (39 ± 1 kg, n = 10). After induction of ventricular fibrillation, biphasic shocks were delivered between two cutaneous patch electrodes (sternal and apical position) or between an esophageal and two cutaneous patch electrodes in a sternal and apical position. The esophageal electrode was integrated into a latex sheath covering a standard transesophageal echocardiography probe.
In 5 of 12 pigs with high body weight, external defibrillation failed despite 3 consecutive 200-J shocks, whereas subsequent transesophageal defibrillation was successful with the first shock. In the remaining 7 pigs, a more than 50% reduction in DFT was obtained with transesophageal defibrillation compared with standard biphasic external defibrillation (67 ± 27 vs 164 ± 23 J,
P < .001). Pigs with lower body weight were successfully defibrillated by both transthoracic and transesophageal shocks. The DFT in pigs with low body weight was significantly lower using transesophageal defibrillation compared with transthoracic shocks (65 ± 15 vs 99 ± 38 J,
P < .05).
In this animal model, nonresponders to standard external defibrillation could successfully be defibrillated via an esophageal-cutaneous electrode configuration. Overall, an almost 50% DFT reduction was achieved by transesophageal defibrillation. Transesophageal defibrillation may provide an additional tool for terminating VF, which is refractory to external defibrillation, eg, in patients with very high body weight. |
---|---|
ISSN: | 0735-6757 1532-8171 |
DOI: | 10.1016/j.ajem.2005.12.014 |