Loading…
diagnostic tool for determining the quality of accuracy validation. Assessing the method for determination of nitrate in drinking water
Realistic internal validation of a method implies the performance validation experiments under intermediate precision conditions. The validation results can be organized in an X NrxNs (replicatesxruns) data matrix, analysis of which enables assessment of the accuracy of the method. By means of Monte...
Saved in:
Published in: | Analytical and bioanalytical chemistry 2007-01, Vol.387 (2), p.619-625 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Realistic internal validation of a method implies the performance validation experiments under intermediate precision conditions. The validation results can be organized in an X NrxNs (replicatesxruns) data matrix, analysis of which enables assessment of the accuracy of the method. By means of Monte Carlo simulation, uncertainty in the estimates of bias and precision can be assessed. A bivariate plot is presented for assessing whether the uncertainty intervals for the bias (E ± U(E)) and intermediate precision (RSDi ± U(RSDi) are included in prefixed limits (requirements for the method). As a case study, a method for determining the concentration of nitrate in drinking water at the official level set by 98/83/EC Directive is assessed by use of the proposed plot. |
---|---|
ISSN: | 1618-2642 1618-2650 |
DOI: | 10.1007/s00216-006-0941-0 |