Loading…
Simultaneous Multiple Immunoassays in a Compact Disc-Shaped Microfluidic Device Based on Centrifugal Force
We explored the potential of a microfluidic device based on centrifugal force as an immunoassay platform by examining the imprecision of assays carried out with 200 nL of sample. Biotinylated antibodies against alpha-fetoprotein (AFP), interleukin-6 (IL-6), and carcinoembryonic antigen [(CEA); 0.1 g...
Saved in:
Published in: | Clinical chemistry (Baltimore, Md.) Md.), 2005-10, Vol.51 (10), p.1955-1961 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We explored the potential of a microfluidic device based on centrifugal force as an immunoassay platform by examining the imprecision of assays carried out with 200 nL of sample.
Biotinylated antibodies against alpha-fetoprotein (AFP), interleukin-6 (IL-6), and carcinoembryonic antigen [(CEA); 0.1 g/L each in 15 mmol/L phosphate-buffered saline (PBS) containing 0.1 mL/L Tween 20] were attached to a microcolumn packed with streptavidin-coated particles. A 200-nL sample was then allowed to pass through the microcolumn for 240 s, followed by Alexa 647-labeled detection antibody (7.5 mg/L in 15 mmol/L PBS containing 10 g/L bovine serum albumin). The flow rate was controlled by altering the rotational speed. Up to 104 sandwich type immunoassays were completed within 50 min.
For AFP, IL-6, and CEA the detection limits were, respectively, 0.15, 1.25, and 1.31 pmol/L. Inter- and intraassay imprecisions (CVs) were |
---|---|
ISSN: | 0009-9147 1530-8561 |
DOI: | 10.1373/clinchem.2005.053348 |