Adaptive signal enhancement of somatosensory evoked potential for spinal cord compression detection: an experimental study

The objective of this study was to assess the efficacy of adaptive signal enhancement (ASE) as a means of indicating intraoperative spinal cord impingement. ASE technique was used to determine the changes in the somatosensory evoked potential (SEP) elicited from eighteen rats with varying levels of...

Full description

Saved in:
Bibliographic Details
Published in:Computers in biology and medicine 2005-11, Vol.35 (9), p.814-828
Main Authors: Hu, Yong, Lam, Benny S.C., Chang, C.Q., Chan, Francis H.Y., Lu, W.W., Luk, Keith D.K.
Format: Article
Language:English
Subjects:
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by
cites
container_end_page 828
container_issue 9
container_start_page 814
container_title Computers in biology and medicine
container_volume 35
creator Hu, Yong
Lam, Benny S.C.
Chang, C.Q.
Chan, Francis H.Y.
Lu, W.W.
Luk, Keith D.K.
description The objective of this study was to assess the efficacy of adaptive signal enhancement (ASE) as a means of indicating intraoperative spinal cord impingement. ASE technique was used to determine the changes in the somatosensory evoked potential (SEP) elicited from eighteen rats with varying levels of spinal cord compression. ASE technique was found to be able to effectively extract SEP signals for the detection of spinal cord injury. Furthermore, while the traditional ensemble averaging (EA) technique requires more than 500 trials for meaningful signal processing in severe noisy SEP recordings, the ASE method required only 50 trials to provide similar information. Because of its fast and reliable SEP detection, the ASE method is ideal for spinal cord monitoring in the clinical setting.
doi_str_mv 10.1016/j.compbiomed.2004.07.002
format article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_68779173</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0010482504000940</els_id><sourcerecordid>68779173</sourcerecordid><originalsourceid>FETCH-LOGICAL-e299t-a759b90d1fb0bc034e4d2485b92d5e5e98e70b7e59f4cdcf23e81a837890f2813</originalsourceid><addsrcrecordid>eNqFkcFu3CAURVGUqJkm_YWKVXd2H9geILskStJKI3XTrBGG55TJ2LjAjDr5-mAlVZddgcR5V7x7CKEMagZs_XVb2zDOvQ8jupoDtDWIGoCfkBWTQlXQNe0pWQEwqFrJu3PyMaUtFBAa-EDO2ZoLyRisyMu1M3P2B6TJP01mR3H6ZSaLI06ZhoGmMJocEk4pxCPFQ3hGR-eQy7Mv9BAiTbNfBm2Iji7fipiSDxN1mNHmcruiZqL4Z8bol9jCprx3x0tyNphdwk_v5wV5vL_7efut2vx4-H57vamQK5UrIzrVK3Bs6KG30LTYOt7KrlfcddihkiigF9ipobXODrxByYxshFQwcMmaC_LlLXeO4fceU9ajTxZ3OzNh2Ce9lkIoJpr_ghy4UKJbEj-_g_u-GNBzWczEo_5bawFu3gAsex08Rp2sx1Kr87F0ol3wmoFeXOqt_udSLy41CF1cNq_TR5dE</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>20279751</pqid></control><display><type>article</type><title>Adaptive signal enhancement of somatosensory evoked potential for spinal cord compression detection: an experimental study</title><source>Elsevier</source><creator>Hu, Yong ; Lam, Benny S.C. ; Chang, C.Q. ; Chan, Francis H.Y. ; Lu, W.W. ; Luk, Keith D.K.</creator><creatorcontrib>Hu, Yong ; Lam, Benny S.C. ; Chang, C.Q. ; Chan, Francis H.Y. ; Lu, W.W. ; Luk, Keith D.K.</creatorcontrib><description>The objective of this study was to assess the efficacy of adaptive signal enhancement (ASE) as a means of indicating intraoperative spinal cord impingement. ASE technique was used to determine the changes in the somatosensory evoked potential (SEP) elicited from eighteen rats with varying levels of spinal cord compression. ASE technique was found to be able to effectively extract SEP signals for the detection of spinal cord injury. Furthermore, while the traditional ensemble averaging (EA) technique requires more than 500 trials for meaningful signal processing in severe noisy SEP recordings, the ASE method required only 50 trials to provide similar information. Because of its fast and reliable SEP detection, the ASE method is ideal for spinal cord monitoring in the clinical setting.</description><identifier>ISSN: 0010-4825</identifier><identifier>EISSN: 1879-0534</identifier><identifier>DOI: 10.1016/j.compbiomed.2004.07.002</identifier><identifier>PMID: 16278110</identifier><language>eng</language><publisher>United States: Elsevier Ltd</publisher><subject>Adaptive signal enhancer ; Algorithms ; Animals ; Ensemble averaging ; Evoked Potentials, Somatosensory ; Intraoperative spinal cord monitoring ; Rats ; Somatosensory evoked potential ; Spinal Cord Compression - diagnosis ; Spinal Cord Compression - physiopathology ; Spinal cord injury</subject><ispartof>Computers in biology and medicine, 2005-11, Vol.35 (9), p.814-828</ispartof><rights>2004 Elsevier Ltd</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/16278110$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Hu, Yong</creatorcontrib><creatorcontrib>Lam, Benny S.C.</creatorcontrib><creatorcontrib>Chang, C.Q.</creatorcontrib><creatorcontrib>Chan, Francis H.Y.</creatorcontrib><creatorcontrib>Lu, W.W.</creatorcontrib><creatorcontrib>Luk, Keith D.K.</creatorcontrib><title>Adaptive signal enhancement of somatosensory evoked potential for spinal cord compression detection: an experimental study</title><title>Computers in biology and medicine</title><addtitle>Comput Biol Med</addtitle><description>The objective of this study was to assess the efficacy of adaptive signal enhancement (ASE) as a means of indicating intraoperative spinal cord impingement. ASE technique was used to determine the changes in the somatosensory evoked potential (SEP) elicited from eighteen rats with varying levels of spinal cord compression. ASE technique was found to be able to effectively extract SEP signals for the detection of spinal cord injury. Furthermore, while the traditional ensemble averaging (EA) technique requires more than 500 trials for meaningful signal processing in severe noisy SEP recordings, the ASE method required only 50 trials to provide similar information. Because of its fast and reliable SEP detection, the ASE method is ideal for spinal cord monitoring in the clinical setting.</description><subject>Adaptive signal enhancer</subject><subject>Algorithms</subject><subject>Animals</subject><subject>Ensemble averaging</subject><subject>Evoked Potentials, Somatosensory</subject><subject>Intraoperative spinal cord monitoring</subject><subject>Rats</subject><subject>Somatosensory evoked potential</subject><subject>Spinal Cord Compression - diagnosis</subject><subject>Spinal Cord Compression - physiopathology</subject><subject>Spinal cord injury</subject><issn>0010-4825</issn><issn>1879-0534</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><recordid>eNqFkcFu3CAURVGUqJkm_YWKVXd2H9geILskStJKI3XTrBGG55TJ2LjAjDr5-mAlVZddgcR5V7x7CKEMagZs_XVb2zDOvQ8jupoDtDWIGoCfkBWTQlXQNe0pWQEwqFrJu3PyMaUtFBAa-EDO2ZoLyRisyMu1M3P2B6TJP01mR3H6ZSaLI06ZhoGmMJocEk4pxCPFQ3hGR-eQy7Mv9BAiTbNfBm2Iji7fipiSDxN1mNHmcruiZqL4Z8bol9jCprx3x0tyNphdwk_v5wV5vL_7efut2vx4-H57vamQK5UrIzrVK3Bs6KG30LTYOt7KrlfcddihkiigF9ipobXODrxByYxshFQwcMmaC_LlLXeO4fceU9ajTxZ3OzNh2Ce9lkIoJpr_ghy4UKJbEj-_g_u-GNBzWczEo_5bawFu3gAsex08Rp2sx1Kr87F0ol3wmoFeXOqt_udSLy41CF1cNq_TR5dE</recordid><startdate>20051101</startdate><enddate>20051101</enddate><creator>Hu, Yong</creator><creator>Lam, Benny S.C.</creator><creator>Chang, C.Q.</creator><creator>Chan, Francis H.Y.</creator><creator>Lu, W.W.</creator><creator>Luk, Keith D.K.</creator><general>Elsevier Ltd</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7QO</scope><scope>8FD</scope><scope>FR3</scope><scope>P64</scope><scope>7X8</scope></search><sort><creationdate>20051101</creationdate><title>Adaptive signal enhancement of somatosensory evoked potential for spinal cord compression detection: an experimental study</title><author>Hu, Yong ; Lam, Benny S.C. ; Chang, C.Q. ; Chan, Francis H.Y. ; Lu, W.W. ; Luk, Keith D.K.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-e299t-a759b90d1fb0bc034e4d2485b92d5e5e98e70b7e59f4cdcf23e81a837890f2813</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>Adaptive signal enhancer</topic><topic>Algorithms</topic><topic>Animals</topic><topic>Ensemble averaging</topic><topic>Evoked Potentials, Somatosensory</topic><topic>Intraoperative spinal cord monitoring</topic><topic>Rats</topic><topic>Somatosensory evoked potential</topic><topic>Spinal Cord Compression - diagnosis</topic><topic>Spinal Cord Compression - physiopathology</topic><topic>Spinal cord injury</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Hu, Yong</creatorcontrib><creatorcontrib>Lam, Benny S.C.</creatorcontrib><creatorcontrib>Chang, C.Q.</creatorcontrib><creatorcontrib>Chan, Francis H.Y.</creatorcontrib><creatorcontrib>Lu, W.W.</creatorcontrib><creatorcontrib>Luk, Keith D.K.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>Biotechnology Research Abstracts</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>MEDLINE - Academic</collection><jtitle>Computers in biology and medicine</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Hu, Yong</au><au>Lam, Benny S.C.</au><au>Chang, C.Q.</au><au>Chan, Francis H.Y.</au><au>Lu, W.W.</au><au>Luk, Keith D.K.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Adaptive signal enhancement of somatosensory evoked potential for spinal cord compression detection: an experimental study</atitle><jtitle>Computers in biology and medicine</jtitle><addtitle>Comput Biol Med</addtitle><date>2005-11-01</date><risdate>2005</risdate><volume>35</volume><issue>9</issue><spage>814</spage><epage>828</epage><pages>814-828</pages><issn>0010-4825</issn><eissn>1879-0534</eissn><abstract>The objective of this study was to assess the efficacy of adaptive signal enhancement (ASE) as a means of indicating intraoperative spinal cord impingement. ASE technique was used to determine the changes in the somatosensory evoked potential (SEP) elicited from eighteen rats with varying levels of spinal cord compression. ASE technique was found to be able to effectively extract SEP signals for the detection of spinal cord injury. Furthermore, while the traditional ensemble averaging (EA) technique requires more than 500 trials for meaningful signal processing in severe noisy SEP recordings, the ASE method required only 50 trials to provide similar information. Because of its fast and reliable SEP detection, the ASE method is ideal for spinal cord monitoring in the clinical setting.</abstract><cop>United States</cop><pub>Elsevier Ltd</pub><pmid>16278110</pmid><doi>10.1016/j.compbiomed.2004.07.002</doi><tpages>15</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0010-4825
ispartof Computers in biology and medicine, 2005-11, Vol.35 (9), p.814-828
issn 0010-4825
1879-0534
language eng
recordid cdi_proquest_miscellaneous_68779173
source Elsevier
subjects Adaptive signal enhancer
Algorithms
Animals
Ensemble averaging
Evoked Potentials, Somatosensory
Intraoperative spinal cord monitoring
Rats
Somatosensory evoked potential
Spinal Cord Compression - diagnosis
Spinal Cord Compression - physiopathology
Spinal cord injury
title Adaptive signal enhancement of somatosensory evoked potential for spinal cord compression detection: an experimental study
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-08T20%3A50%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Adaptive%20signal%20enhancement%20of%20somatosensory%20evoked%20potential%20for%20spinal%20cord%20compression%20detection:%20an%20experimental%20study&rft.jtitle=Computers%20in%20biology%20and%20medicine&rft.au=Hu,%20Yong&rft.date=2005-11-01&rft.volume=35&rft.issue=9&rft.spage=814&rft.epage=828&rft.pages=814-828&rft.issn=0010-4825&rft.eissn=1879-0534&rft_id=info:doi/10.1016/j.compbiomed.2004.07.002&rft_dat=%3Cproquest_pubme%3E68779173%3C/proquest_pubme%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-e299t-a759b90d1fb0bc034e4d2485b92d5e5e98e70b7e59f4cdcf23e81a837890f2813%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=20279751&rft_id=info:pmid/16278110&rfr_iscdi=true