Loading…

Thrombus formation on an atrial septal defect closure device: a case report and review of the literature

We report on a case of a mobile left atrial thrombus formation on an atrial septal defect occluder system (28 mm StarFLEX-Occluder) despite 6 months of postprocedural anticoagulation with phenprocoumon and platelet antiaggregation with aspirin in a 69-year-old woman. The closure was performed becaus...

Full description

Saved in:
Bibliographic Details
Published in:European journal of echocardiography 2007-01, Vol.8 (1), p.53-56
Main Authors: Divchev, Dimitar, Schaefer, Arnd, Fuchs, Martin, Breymann, Thomas, Drexler, Helmut, Meyer, Gerd Peter
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:We report on a case of a mobile left atrial thrombus formation on an atrial septal defect occluder system (28 mm StarFLEX-Occluder) despite 6 months of postprocedural anticoagulation with phenprocoumon and platelet antiaggregation with aspirin in a 69-year-old woman. The closure was performed because of a significant left to right atrial shunt (Qp/Qs 1.8) with enlargement of the right atrial and ventricular cavities and impairment of right ventricular function in the presence of persistent atrial fibrillation and chronic heart failure (NYHA II-III). The 6-month follow up by transoesophageal echocardiography (TEE) revealed the floating thrombus located at the left atrial side of the occluder.
ISSN:1525-2167
DOI:10.1016/j.euje.2005.10.011