Loading…

A routine high-performance size-exclusion chromatography to determine molecular size distribution of Haemophilus influenzae type b conjugate vaccines

High-performance size exclusion chromatography has been used to determine the molecular size distribution of Haemophilus influenzae type b (Hib) conjugate vaccines. Both high molecular weight preparations of native Hib capsular polysaccharide coupled to tetanus toxoid and low molecular weight vaccin...

Full description

Saved in:
Bibliographic Details
Published in:Vaccine 1999, Vol.17 (2), p.118-125
Main Authors: von Hunolstein, C, Parisi, L, Recchia, S
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:High-performance size exclusion chromatography has been used to determine the molecular size distribution of Haemophilus influenzae type b (Hib) conjugate vaccines. Both high molecular weight preparations of native Hib capsular polysaccharide coupled to tetanus toxoid and low molecular weight vaccines with Hib oligosaccharides linked to the CRM 197 nontoxic mutant diphtheria protein were analysed. Columns with different fractionation ranges were used for the two kinds of vaccines. This method showed to be rapid, accurate and reproducible for different lots of Hib vaccine of different composition produced by various manufacturers. It could replace more time-consuming chromatographic methods enabling control authorities to employ a single methodological approach for different Hib vaccines.
ISSN:0264-410X
1873-2518