Utility of multislice computed tomography with a 64-data acquisition system for four-dimensional volumetric analysis using a pulsating phantom and considering pulsation rate and reconstruction methods

Abstract Purpose To evaluate 64-data acquisition system (DAS) MSCT (Light Speed VCT, GE) at 0.625 mm slice thickness, 0.35 s/rotation, tube 120 kV at 400 mA, ECG-gated for 4-D volumetric analysis, we used pulsating phantoms to measure end-diastolic (EDV) and end-systolic (ESV) volume and ejection fr...

Full description

Saved in:
Bibliographic Details
Published in:International journal of cardiology 2008-11, Vol.130 (3), p.477-480
Main Authors: Mizuno, Naoko, Funabashi, Nobusada, Nakamura, Koki, Uehara, Masae, Suzuki, Kazushi, Terao, Makoto, Okubo, Kenji, Mita, Yuzuru, Maeda, Fumiaki, Komuro, Issei
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract Purpose To evaluate 64-data acquisition system (DAS) MSCT (Light Speed VCT, GE) at 0.625 mm slice thickness, 0.35 s/rotation, tube 120 kV at 400 mA, ECG-gated for 4-D volumetric analysis, we used pulsating phantoms to measure end-diastolic (EDV) and end-systolic (ESV) volume and ejection fraction (EF) to assess reconstruction methods especially for higher pulsation rates. Materials and methods A pulsating device (AZ-631N, Anzai Medical) with contrast material (300 mgI/dl) diluted 10× with saline was moved at 40–110 to-and-fro movements/min. ECG-gated MSCT was performed × 5 per pulsation rate. The EDV and ESV were measured using workstation (Virtual Place Advance Plus, Aze). Results The mean EDV and ESV were 98, 97, 97 96, 95, 94, and 101% and 145, 143, 142, 144, 145, 149, 156 and 160%, respectively, compared to the static state. EF was 80, 81, 81, 80, 79, 77, 73, and 76% at 40–110 pulsations/min, when reconstructed by the segmented method, but was improved to 82, 83, 85, and 84% at 80–110 beat/min when reconstructed by the burst method. The latter is therefore more appropriate for higher rates. Conclusion This 64-DAS MSCT can measure EDV even at high beat rates (up to 110 beats per minute) compared to the static state. Because ESV tended to be overestimated by approximately 140–160% compared with the static state, EF tended to be underestimated by approximately 73–81% compared with the static state. However, at higher beat rates of > 70 beat/min, an appropriate reconstruction method (the burst method) may further improve the accuracy of EF measurement.