We report a case of palatopharyngeal wall perforation during intubation with a GlideScope laryngoscope. The likely mechanism was advancing and rotating the endotracheal tube against a taut palatopharyngeal fold. This was missed during the initial laryngoscopy, because there is a potential blind-spot...
Saved in:
Published in: | Anaesthesia and intensive care 2008-11, Vol.36 (6), p.870-874 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We report a case of palatopharyngeal wall perforation during intubation with a GlideScope laryngoscope. The likely mechanism was advancing and rotating the endotracheal tube against a taut palatopharyngeal fold. This was missed during the initial laryngoscopy, because there is a potential blind-spot in the oropharynx when attention is focused on the GlideScope" monitor Fortunately, there were no sequelae other than minor bleeding and a mild sore throat and no surgical intervention was necessary. The use of unnecessary force during the endotracheal tube insertion, the use of too large a laryngoscope blade and the use of a rigid stylet could possibly also have been contributory factors to this complication. |
---|---|