Stenting of long coronary artery lesions: Initial angiographic results and 6-month clinical outcome of the micro stent II-XL
To evaluate the results of long Micro Stent II (MS‐XL) implantations, 119 MS‐XLs were implanted in 102 patients (age, 62.83 years). Nineteen stents (16%) were implanted in saphenous vein grafts; 100 stents (84%) were implanted in native coronary arteries. Twenty‐five patients (25%) were treated beca...
Saved in:
Published in: | Catheterization and cardiovascular interventions 1999-09, Vol.48 (1), p.105-112 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To evaluate the results of long Micro Stent II (MS‐XL) implantations, 119 MS‐XLs were implanted in 102 patients (age, 62.83 years). Nineteen stents (16%) were implanted in saphenous vein grafts; 100 stents (84%) were implanted in native coronary arteries. Twenty‐five patients (25%) were treated because of acute myocardial infarction (AMI); 30 patients (29%) because of unstable angina or angina class IV, and 47 patients (46%) because of stable angina. Eighty‐six de novo lesions (84%) and 16 restenotic lesions (16%) were treated. Indications for stent implantation include elective, 61 patients (60%); suboptimal balloon angioplasty result, 22 patients (21%); and bailout after balloon angioplasty, 19 patients (19%). Because of residual thrombus after stenting, 27 patients (26%) received abciximab. All patients received ticlopidin for 28 days and acetylsalicylic acid. One hundred and seventeen MS‐XLs (98%) were implanted successfully. Additional (shorter) MS‐II were implanted in 40 patients (39%). The stented segment length was 45 ± 20 mm. The minimum lumen diameter increased from 0.5 ± 0.5 mm before to 2.7 ± 0.5 mm after stent implantation. The acute gain was 2.2 ± 0.4 mm. Early clinical events ( |
---|---|
ISSN: | 1522-1946 1522-726X |
DOI: | 10.1002/(SICI)1522-726X(199909)48:1<105::AID-CCD22>3.0.CO;2-0 |