The reliability of the i-STAT clinical portable analyser

Summary The purpose of this investigation was to assess the reliability of the i-STAT clinical portable analyser and CG4+ cartridge measures of blood pH, partial pressures of O2 ( p O2 ) and CO2 ( p CO2 ), bicarbonate ([HCO3− ]), base excess (BE), total carbon dioxide ( T CO2 ), oxygen saturation (...

Full description

Saved in:
Bibliographic Details
Published in:Journal of science and medicine in sport 2007-06, Vol.10 (3), p.135-140
Main Authors: Dascombe, B.J, Reaburn, P.R.J, Sirotic, A.C, Coutts, A.J
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Summary The purpose of this investigation was to assess the reliability of the i-STAT clinical portable analyser and CG4+ cartridge measures of blood pH, partial pressures of O2 ( p O2 ) and CO2 ( p CO2 ), bicarbonate ([HCO3− ]), base excess (BE), total carbon dioxide ( T CO2 ), oxygen saturation ( s O2 ) and blood lactate ([BLa− ]) at various exercise intensities. A comparison between [BLa− ] measured by the i-STAT and the Accusport lactate analysers during an intermittent treadmill run was also undertaken. The technical error of measurement (TEM%) at rest, at moderate (200 W) and maximal exercise ( V ˙ O 2   max ) was acceptable (