Adverse events with transvenous left ventricular pacing in patients with severe heart failure: early experience from a single centre
Assessment of complications following implantation of transvenous ventricular electrodes to pace the left ventricle. Twenty-eight patients with severe cardiac failure and left bundle branch block were prospectively followed for adverse effects of implantation of a left ventricular transvenous pacing...
Saved in:
Published in: | Europace (London, England) England), 2001-01, Vol.3 (1), p.60-63 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Assessment of complications following implantation of transvenous ventricular electrodes to pace the left ventricle.
Twenty-eight patients with severe cardiac failure and left bundle branch block were prospectively followed for adverse effects of implantation of a left ventricular transvenous pacing system. Immediate follow-up was associated with loss of left ventricular pacing in nine patients (32%). This was due to lead dislodgement in four cases (corrected by re-operation in three of these cases), and due to increased threshold in five cases (corrected by programming a higher pacing amplitude in all five cases, but with intermittent diaphragmatic contraction in one case). After 1 month, one patient died, one patient with severe coronary heart disease suffered a myocardial infarction, and left ventricular pacing was lost in two patients. Pericardial effusion, new significant ventricular arrhythmias or other adverse effects were not observed. After a mean follow-up of 16 +/- 9.2 months, pacing leads remained stable and no late complications related to the transvenous left ventricular epicardial pacing were observed.
Placement of a permanent lead in a tributary of the coronary sinus is feasible without serious adverse effects during the first month. The only frequent adverse event was lead dislodgement; a finding which emphasizes the need for development of specially designed leads for this application. |
---|---|
ISSN: | 1099-5129 |
DOI: | 10.1053/eupc.2000.0138 |