Loading…
Miniaturized solid-phase extraction with resin disks
The properties and advantages of membrane disks for solid-phase extraction (SPE) are described. Miniaturization is a trend in SPE, as well as chemical analysis in general. A semimicro method is reviewed in which an extraction disk 4 mm in diameter is used for SPE. Even smaller scale separations are...
Saved in:
Published in: | Journal of Chromatography A 2001-02, Vol.909 (1), p.79-85 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The properties and advantages of membrane disks for solid-phase extraction (SPE) are described. Miniaturization is a trend in SPE, as well as chemical analysis in general. A semimicro method is reviewed in which an extraction disk 4 mm in diameter is used for SPE. Even smaller scale separations are possible with a device in which a membrane 0.7 mm in diameter was incorporated into the needle of a 50-μl syringe. Aqueous samples containing 10 ppb of eight substituted benzenes were passed through the miniaturized-SPE syringe automatically using a single-syringe infusion pump. Elution volumes of 5 μl provided 500-fold concentrations and delivered average recoveries greater than 90% and an average relative standard deviation (RSD) of 4.6% for the analytes. Direct injection of the 5 μl eluate from the miniaturized-SPE syringe into a gas chromatograph also produced average recoveries greater than 90% and an average RSD of 6.2%. |
---|---|
ISSN: | 0021-9673 |
DOI: | 10.1016/S0021-9673(00)01014-1 |