Collection of hyperconcentrated platelets with Trima Accel
Background and Objectives Lowering the plasma content in single‐donor platelet (PLT) concentrates well below 30% implies the need to collect platelets at very high concentrations. Trima Accel (TA) is validated for collection below 4000 × 103 PLTs/µl. We evaluated its performance at 5000 × 103 PLTs/...
Saved in:
Published in: | Vox sanguinis 2006-02, Vol.90 (2), p.92-96 |
---|---|
Main Authors: | , , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Background and Objectives Lowering the plasma content in single‐donor platelet (PLT) concentrates well below 30% implies the need to collect platelets at very high concentrations. Trima Accel (TA) is validated for collection below 4000 × 103 PLTs/µl. We evaluated its performance at 5000 × 103 PLTs/µl.
Materials and Methods Twenty blood donors underwent apheresis with TA twice collecting either a hyperconcentrated or a standard single‐donor platelet concentrate with a target platelet concentration of 5000 or 1200 × 103 PLTs/µl, respectively. We analysed the collection efficiency, the collection rate and the quality of the collected by‐plasma.
Results We collected 20 hyperconcentrated and 20 standard units containing 2·56 ± 0·5 and 3·39 ± 0·4 × 1011 PLTs at a concentration of 4518 ± 978 and 1374 ± 166 × 103 PLTs/µl in 45 ± 8 and 39 ± 6 min resulting in a collection efficiency of 47·5 ± 10·0 and 70·7 ± 7·9% and a collection rate of 5·9 ± 1·4 and 8·8 ± 1·5 × 109 PLTs/min, respectively (all results expressed as mean ± standard deviation). The collected by‐plasma showed a very high grade of cell purity and a satisfactory recovery of the clotting factors.
Conclusion Although TA is a suitable device for PLT collection at very high concentrations, improvements are desirable to further increase the productivity above its currently validated upper collect concentration limit. |
---|---|