Loading…
Washing Machine Associated 50 Hz Detected As Ventricular Fibrillation by An Implanted Cardioverter Defibrillator
SABATÉ, X., et al.: Washing Machine Associated 50 Hz Detected As Ventricular Fibrillation by An Implanted Cardioverter Defibrillator. This case report describes a patient with an automatic ICD who suffered a defibrillation shock without warning symptoms. An electrical interference can be observed in...
Saved in:
Published in: | Pacing and clinical electrophysiology 2001-08, Vol.24 (8), p.1281-1283 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | SABATÉ, X., et al.: Washing Machine Associated 50 Hz Detected As Ventricular Fibrillation by An Implanted Cardioverter Defibrillator. This case report describes a patient with an automatic ICD who suffered a defibrillation shock without warning symptoms. An electrical interference can be observed in the stored EGM of the episode. The patient explained that the moment he felt the shock he was touching a washing machine. After correct grounding of this machine the patient did not suffer more inappropriate shocks. |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1046/j.1460-9592.2001.01281.x |