Validation of a novel method for continuous saline tonometry in a porcine model
Only intermittent and semi-continuous tonometric measurement of gastric and intestinal pHi is possible with the equipment available today. Earlier we developed a system for continuous saline tonometry and tested it in vitro. To assess the in vivo reliability of this method for continuous gastrointes...
Saved in:
Published in: | Physiological measurement 2001-08, Vol.22 (3), p.453-460 |
---|---|
Main Authors: | , , , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | Only intermittent and semi-continuous tonometric measurement of gastric and intestinal pHi is possible with the equipment available today. Earlier we developed a system for continuous saline tonometry and tested it in vitro. To assess the in vivo reliability of this method for continuous gastrointestinal saline tonometry, a standard tonometer for measurement of intestinal pCO2 and corresponding pHi was modified to allow continuous perfusion of physiological saline in a closed system and tested in a porcine model. In 11 anaesthetized and haemodynamically stable pigs, two continuous tonometry balloons were inserted into the distal small bowel, and a standard tonometry balloon was used as reference. To test long-term function of the continuous tonometers the research protocol lasted for eight hours. The two continuous saline tonometers performed well, and after an equilibration time of three hours the mean pHi values were stable between 7.35 and 7.43 and between 7.32 and 7.39 respectively. The standard tonometer measured stable pHi values. These preliminary studies indicate that continuous saline tonometry performs well over eight hours with a small bias and a good precision. |
---|---|
ISSN: | 0967-3334 1361-6579 1361-6579 |
DOI: | 10.1088/0967-3334/22/3/304 |