Ultrasonic scaler tip performance under various load conditions

Objectives: The aim of this investigation was to assess the effect of loading on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer. Material and Methods: Four generators were selected for this investigation including: Cavitron Select (Dentsply), Cavitron SPS (Dentsply),...

Full description

Saved in:
Bibliographic Details
Published in:Journal of clinical periodontology 2003-10, Vol.30 (10), p.876-881
Main Authors: Lea, S. C., Landini, G., Walmsley, A. D.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Objectives: The aim of this investigation was to assess the effect of loading on ultrasonic scaler tip displacement amplitude, using a scanning laser vibrometer. Material and Methods: Four generators were selected for this investigation including: Cavitron Select (Dentsply), Cavitron SPS (Dentsply), Piezon Master 400 (EMS) and Mini Piezon (EMS). The insert designs used with these generators included the TFI‐10 and TFI‐3 (with the Dentsply generators) and P‐tip (with the EMS generators). The insert tips were contacted against tooth surfaces with loads of 0.25, 0.5 and 1.0 N. The scanning laser vibrometer was then used to scan the vibrating scalers at various generator power settings. Results: The tips generally showed an increase in displacement amplitude with increasing power setting, although this increase was often nonlinear. Displacement amplitudes of like tips were generally found to be significantly different from each other at all loads. For individual inserts, loading caused a significant difference in tip response from the unloaded situation and also between loads. Conclusions: This investigation has demonstrated that there is a high variability associated with dental ultrasonic scaler inserts. Clinicians and researchers should be aware of this variability as it may affect both clinical procedures and future research. Zusammenfassung Die Bewährung von Ultraschallscaleransätzen unter verschiedenen Belastungen Zielsetzung:  Untersuchung des Einflusses der Belastung auf die Bewegungsamplitude von Ultraschallscalerspitzen mittels eines Scanning‐Laser‐Vibrometers (SLV). Material und Methoden:  4 Ultraschallgeräte wurden für diese Studie ausgewählt: Cavitron Select (Dentsply), Cavitron SPS (Dentsply), Piezon Master 400 (EMS) und Mini Piezon (EMS). Die Ansatzformen, die mit diesen Geräten benutzt wurden, umfassten den TFI‐10 und TFI‐3 (mit den Dentsply‐Geräten) sowie die P‐Spitze (mit den EMS‐Geräten). Die Ansatzspitzen wurden mit Kräften von 0.25, 0.5 und 1.0 N mit Zahnoberflächen in Kontakt gebracht. Das SLV wurde dann benutzt, um die schwingenden Scaler bei unterschiedlichen Energieeinstellungen der Geräte aufzunehmen. Ergebnisse:  Die Spitzen zeigten generell eine Zunahme der Bewegungsamplitude mit zunehmender Energie. Allerdings war dieser Anstieg häufig nicht linear. Die Bewegungsamplituden gleichartiger Spitzen unterschieden sich signifikant voneinander bei allen Belastungen. Für gleiche Ansätze resultierte Belastung in einem signifikanten Untersch
ISSN:0303-6979
1600-051X
DOI:10.1034/j.1600-051X.2003.00395.x