Loading…
Evaluation of a new handheld biosensor for point-of-care testing of whole blood beta-hydroxybutyrate concentration
To evaluate performance characteristics of the newly available handheld combined glucose and ketone meter for beta-hydroxybutyrate measurement. Laboratory method evaluation. Accuracy of beta-hydroxybutyrate measurement and effect of acetoacetate interference at clinically important beta-hydroxybutyr...
Saved in:
Published in: | Hong Kong medical journal = Xianggang yi xue za zhi 2002-06, Vol.8 (3), p.172-176 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | To evaluate performance characteristics of the newly available handheld combined glucose and ketone meter for beta-hydroxybutyrate measurement.
Laboratory method evaluation.
Accuracy of beta-hydroxybutyrate measurement and effect of acetoacetate interference at clinically important beta-hydroxybutyrate levels.
Deming regression analysis of beta-hydroxybutyrate measurements assessed by the ketone sensor and a laboratory enzymatic method revealed a coefficient of determination of 0.989 (P |
---|---|
ISSN: | 1024-2708 2226-8707 |