Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia

KRISHEN, A., et al.: Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia. This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glu...

Full description

Saved in:
Bibliographic Details
Published in:Pacing and clinical electrophysiology 2001-11, Vol.24 (11), p.1701-1703
Main Authors: KRISHEN, AJAY, SHEPARD, RICHARD K., LEFFLER, JEANINE A., WOOD, MARK A., ELLENBOGEN, KENNETH A.
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:KRISHEN, A., et al.: Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia. This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glucose levels and this finding was reproduced with a glucose challenge. No T wave oversensing was seen during treatment of hyperglycemia. Alterations in serum chemistry may account for intermittent T wave oversensing in patients with ICDs.