Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia
KRISHEN, A., et al.: Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia. This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glu...
Saved in:
Published in: | Pacing and clinical electrophysiology 2001-11, Vol.24 (11), p.1701-1703 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | KRISHEN, A., et al.: Implantable Cardioverter Defibrillator T Wave Oversensing Caused by Hyperglycemia. This case report highlights a previously unreported cause of T wave oversensing in a patient with an ICD and recent onset diabetes. Consistent T wave oversensing was observed at elevated serum glucose levels and this finding was reproduced with a glucose challenge. No T wave oversensing was seen during treatment of hyperglycemia. Alterations in serum chemistry may account for intermittent T wave oversensing in patients with ICDs. |
---|---|