In vivo and in vitro precision for bone density measured by dual-energy X-ray absorption

An investigation was made into some of the major sources of error influencing the bone mineral density (BMD) measurements of the lumbar vertebrae, the femoral neck and the greater trochanter. The effect on accuracy and reproducibility of the following parameters was investigated: influence of patien...

Full description

Saved in:
Bibliographic Details
Published in:Osteoporosis international 1991-06, Vol.1 (3), p.141-146
Main Authors: Lilley, J, Walters, B G, Heath, D A, Drolc, Z
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:An investigation was made into some of the major sources of error influencing the bone mineral density (BMD) measurements of the lumbar vertebrae, the femoral neck and the greater trochanter. The effect on accuracy and reproducibility of the following parameters was investigated: influence of patient positioning, patient size, scan speed, the technique of scan analysis and the temporal variation in instrument performance. The in vitro precision, both long-term and short-term, was assessed using aluminium phantoms supplied by the manufacturer. For the spine phantom, the precision expressed as a percentage coefficient of variation (%CV) was found to be 0.4% (10 scans) in the short term and 0.55% (15 scans) in the long term. Measured precision (short-term) for the three regions of the femur phantom analysed by the software was 1.3% for the neck of femur, 1.7% for Ward's triangle and 0.6% for the trochanter. Long-term precision was 1.0%, 1.9% and 1.1% respectively. No statistically significant difference was found between long- and short-term results. Short-term in vitro precision on a low density anthropological phantom was 4.1%, 4.2%, 2.4% and 0.61% for neck of femur, Ward's triangle, trochanter and spine respectively. In vivo short-term precision for the lumbar spine (L2-L4), measured by scanning four normal volunteers five times in one session, was found to be 0.8 +/- 0.25%. In vivo precision for the femur, measured on seven volunteers was 1.6 +/- 0.8% for the neck of femur, 3.2 +/- 1.7% for Ward's triangle and 2.2 +/- 1.1% for the trochanter.
ISSN:0937-941X
1433-2965
DOI:10.1007/BF01625443