Loading…

Element release from titanium devices used in oral and maxillofacial surgery

Optical microscopy, scanning electron microscopy and X-ray microanalysis (EDS system) were used on c.p. titanium devices (21 grids and 10 plates) removed from 28 patients without signs of inflammation 6–24 months after surgery. Plates, grids and surrounding tissue were investigated to evaluate the t...

Full description

Saved in:
Bibliographic Details
Published in:Biomaterials 2003-03, Vol.24 (6), p.1093-1099
Main Authors: Zaffe, Davide, Bertoldi, Carlo, Consolo, Ugo
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Optical microscopy, scanning electron microscopy and X-ray microanalysis (EDS system) were used on c.p. titanium devices (21 grids and 10 plates) removed from 28 patients without signs of inflammation 6–24 months after surgery. Plates, grids and surrounding tissue were investigated to evaluate the titanium release and accumulation. Titanium was only present in the interfacial bone, probably due to fretting, and in all fibrous tissue surrounding the devices. Titanium content followed a decreasing gradient extending from the device surface and was not detected at a distance greater than 1 mm. High titanium levels were found in blood cells in the connective tissue. In conclusion, titanium release from the devices stops only after bone is laid down on the titanium surfaces. Titanium release does not seem to interfere with the osteogenic process but perhaps may interact with it.
ISSN:0142-9612
1878-5905
DOI:10.1016/S0142-9612(02)00424-6