Identification and quantification of metallothionein isoforms and superoxide dismutase in spiked liver extracts using HPLC-ESI-MS offline coupling and HPLC-ICP-MS online coupling
A two-dimensional chromatographic method for the characterization of metallothionein isoforms (MT) and superoxide dismutase (SOD) in spiked liver extracts was developed for the optimization of extraction procedures from liver samples. Element-specific detection (ICP-MS) and molecule-specific detecti...
Saved in:
Published in: | Analytical and bioanalytical chemistry 2003-01, Vol.375 (1), p.145-156 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A two-dimensional chromatographic method for the characterization of metallothionein isoforms (MT) and superoxide dismutase (SOD) in spiked liver extracts was developed for the optimization of extraction procedures from liver samples. Element-specific detection (ICP-MS) and molecule-specific detection (ESI-MS) were applied for maximum species information. A special focus was laid on the quantitative data evaluation (species stoichiometry, calibration with and without matrix, recovery), which is neglected in most MT/SOD publications with hyphenated techniques. Linearity, precision (residual standard deviation of calibration curves |
---|---|
ISSN: | 1618-2642 |
DOI: | 10.1007/s00216-002-1594-2 |