A simple device to inject indicator gas for wash-out tests during mechanical ventilation

To evaluate a simple device which injects a constant fraction of indicator gas to the inspiratory mixture for performing multi-breath wash-out tests during controlled ventilation. the technique in which the indicator gas is injected at the mouth of the patient (post-mix) is compared with the techniq...

Full description

Saved in:
Bibliographic Details
Published in:Intensive care medicine 1992-05, Vol.18 (5), p.304-308
Main Authors: GÜLTUNA, I, HUYGEN, P. E. M, JABAAIJ, C, HOLLAND, W. P. J, INCE, C, BRUINING, H. A
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:To evaluate a simple device which injects a constant fraction of indicator gas to the inspiratory mixture for performing multi-breath wash-out tests during controlled ventilation. the technique in which the indicator gas is injected at the mouth of the patient (post-mix) is compared with the technique where the indicator gas is administered in the bellows of the ventilator (pre-mix). Surgical Intensive Care Unit of a University Hospital. 10 post-operative mechanically ventilated patients. None. 3 wash-out tests with the post-mix and 3 wash-out tests with the pre-mix method were performed within an hour on every patient. The calculated mean end expiratory lung volume (EEV) was 1.91 +/- 0.871 with the post-mix technique and 1.89 +/- 0.881 with the pre-mix technique. There was a good agreement with a mean difference of -1.9 +/- 6.5% in the calculated EEV values by the two different techniques. The described injector is an affordable device, is easy to assemble and can be incorporated in most electronically regulated ventilators to perform multi-breath indicator gas wash-out tests for pulmonary monitoring at the bed side of ICU patients.
ISSN:0342-4642
1432-1238
DOI:10.1007/BF01706480