Hybrid retinal imager using line-scanning laser ophthalmoscopy and spectral domain optical coherence tomography
We demonstrate for the first time the integration of two technologies, Spectral Domain Optical Coherence Tomography (SDOCT) and Line-Scanning Laser Ophthalmoscopy (LSLO) into a single compact instrument that shares the same imaging optics and line scan camera for both OCT and LSLO imaging. Co-regist...
Saved in:
Published in: | Optics express 2006-12, Vol.14 (26), p.12909-12914 |
---|---|
Main Authors: | , , , , , |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | We demonstrate for the first time the integration of two technologies, Spectral Domain Optical Coherence Tomography (SDOCT) and Line-Scanning Laser Ophthalmoscopy (LSLO) into a single compact instrument that shares the same imaging optics and line scan camera for both OCT and LSLO imaging. Co-registered high contrast wide-field en face retinal LSLO and SDOCT images are obtained non-mydriatically with less than 600 microwatts of broadband illumination at 15 frames/sec. The LSLO/SDOCT hybrid instrument could have important applications in clinical ophthalmic diagnostics and emergency medicine. |
---|---|
ISSN: | 1094-4087 1094-4087 |
DOI: | 10.1364/OE.14.012909 |