A simplified liquid chromatography tandem mass spectrometry assay, using on-line solid-phase extraction, for the quantitation of cortisol in saliva and comparison with a routine DELFIA method

Background We have developed a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for measuring salivary cortisol, which requires only 200 μL sample and no extraction. Methods Sample (200 μL) and 25 μL internal standard were added directly to a 96-deep-well plate. Of this, 50 μL was lo...

Full description

Saved in:
Bibliographic Details
Published in:Annals of clinical biochemistry 2010-03, Vol.47 (2), p.131-136
Main Authors: Owen, L J, Haslam, S, Adaway, J E, Wood, P, Glenn, C, Keevil, B G
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Background We have developed a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for measuring salivary cortisol, which requires only 200 μL sample and no extraction. Methods Sample (200 μL) and 25 μL internal standard were added directly to a 96-deep-well plate. Of this, 50 μL was loaded onto a guard cartridge, the cartridge was then washed and the eluate was diverted to waste. The compounds were eluted from the guard cartridge onto the C18 analytical column. Cortisol and deuterated cortisol were monitored using transitions m/z 363.2 > 121.1 and 365.1 > 122.2, respectively. Results The method had a lower limit of quantitation of 2 nmol/L. Intra-assay and inter-assay imprecision were better than 9.5%. Comparison with an established dissociation enhanced lanthanide fluorescent immunoassay (DELFIA) gave good agreement for the majority of samples; LC-MS/MS = 1.0065 × DELFIA − 3.7 (n = 130). The reference range was determined to be 5.8–45.7 nmol/L at 08:00 h and