Hydralazine-selective PVC membrane electrode based on hydralazinium tetraphenylborate

A hydralazine ion-selective PVC membrane electrode based on hydralazinium tetraphenylborate has been prepared with dioctyl phthalate as plasticizer. The electrode showed linear response with a slope factor of 57.5 mV/concentration decade at 20° over the concentration range from 4 × 10 −4 to 10 −1 M...

Full description

Saved in:
Bibliographic Details
Published in:Talanta (Oxford) 1988-06, Vol.35 (6), p.487-489
Main Authors: Badawy, S.S., Shoukry, A.F., Rizk, M.S., Omar, M.M.
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:A hydralazine ion-selective PVC membrane electrode based on hydralazinium tetraphenylborate has been prepared with dioctyl phthalate as plasticizer. The electrode showed linear response with a slope factor of 57.5 mV/concentration decade at 20° over the concentration range from 4 × 10 −4 to 10 −1 M hydralazine. The effects, on the electrode performance, of membrane composition, pH of the test solution and the time of soaking were studied. The electrode exhibited good selectivity for hydralazine with respect to a large number of inorganic cations and organic substances of biological importance. The standard-addition method and potentiometric titrations were used to determine hydralazine concentrations in pure solutions and in a pharmaceutical preparation, with satisfactory results.
ISSN:0039-9140
1873-3573