Loading…
Photoacoustic spectroscopic studies on the solid phase cell with a differential type microphone using a diode laser as radiation source
The design and applications of the solid phase photoacoustic cell with a differential type microphone are described. This cell has higher sensitivity and reproducibility relative to a cell without a differential microphone. The method has been applied to the determination of phosphate ion in water s...
Saved in:
Published in: | Talanta (Oxford) 1994-11, Vol.41 (11), p.1861-1864 |
---|---|
Main Authors: | , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The design and applications of the solid phase photoacoustic cell with a differential type microphone are described. This cell has higher sensitivity and reproducibility relative to a cell without a differential microphone. The method has been applied to the determination of phosphate ion in water samples. The detection limit for phosphate ion concentrated from 22.5 ml of sample solution on a membrane filter as molybdophosphate-
n-dodecyltrimethylammonium bromide was 3.0 ng PO
3−
4/ml and the coefficient of variation for five measurements at 20 ng PO
3−
4/ml of phosphate ion was 3% using a diode laser (30 mW, 826 nm) as radiation source. The calibration graph for phosphate ion was linear over the range 5–50 ng PO
3−
4/ml. |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/0039-9140(94)80192-4 |