Modelling the background current with partial least squares regression and transference of the calibration models in the simultaneous determination of Tl and Pb by stripping voltammetry

With the aim of carrying out a calibration transfer for routine analysis, partial least squares (PLS) regression was successfully applied to simultaneously determine thallium and lead by stripping voltammetry when an interfering background current is present. The presence of a significant blank sign...

Full description

Saved in:
Bibliographic Details
Published in:Talanta (Oxford) 1998-05, Vol.46 (1), p.129-138
Main Authors: Herrero, Ana, Ortiz, M.Cruz
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:With the aim of carrying out a calibration transfer for routine analysis, partial least squares (PLS) regression was successfully applied to simultaneously determine thallium and lead by stripping voltammetry when an interfering background current is present. The presence of a significant blank signal that overlaps the thallium peak, together with the overlapping thallium and lead signals were both suitably modelled by this multivariate regression technique. Moreover, once the PLS models are built, the piecewise direct standardization (PDS) method can be used to transfer these models over time in such a way that the number of calibration samples that will be needed in future determinations is reduced from 25 to 9, without a loss of quality in the analyses. The mean of the relative errors (in absolute values) obtained for thallium and lead is below 4.94% and 3.19%, respectively.
ISSN:0039-9140
1873-3573