Cesium ion selective electrode based on calix[4]crown ether–ester
The potentiometric response characteristics of cesium ion selective PVC membrane electrode employing calix[4]crown ether–ester as an ionophore were investigated. The electrode exhibit a good response for cesium ion over wide concentration range of 5.0×10 −6–1.0×10 −1 M with a Nernstian slope of 59 m...
Saved in:
Published in: | Talanta (Oxford) 2002-09, Vol.58 (3), p.445-450 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The potentiometric response characteristics of cesium ion selective PVC membrane electrode employing calix[4]crown ether–ester as an ionophore were investigated. The electrode exhibit a good response for cesium ion over wide concentration range of 5.0×10
−6–1.0×10
−1 M with a Nernstian slope of 59 mV per decade. The detection limit of electrode is 5.0×10
−6 M. The electrode was found to have selectivity for cesium ion over alkali, alkaline and transition metals. The response time of the electrode is less than 20 s and can be used for more than 4 months without observing any divergence in potentiometric response. The electrode response was stable over wide pH range. |
---|---|