Flow-injection chemiluminescence determination of formaldehyde in water
A modification of the Trautz–Schorigin reaction into a flow-injection analysis configuration is described. Different approaches were used at the optimization of chemiluminescence determination of formaldehyde in water based on the reaction of formaldehyde, gallic acid and hydrogen peroxide in an alk...
Saved in:
Published in: | Talanta (Oxford) 2007-02, Vol.71 (2), p.900-905 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Subjects: | |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A modification of the Trautz–Schorigin reaction into a flow-injection analysis configuration is described. Different approaches were used at the optimization of chemiluminescence determination of formaldehyde in water based on the reaction of formaldehyde, gallic acid and hydrogen peroxide in an alkaline solution. Detection system with a 218
μl chemiluminescence cell was optimized by both a one-variable-at-a-time method, and a modified simplex method. A calibration graph is linear in the concentration range 4
×
10
−8 to 1
×
10
−5
M HCHO. The detection limit of formaldehyde for a signal-to-noise ratio of 3 is 4
×
10
−8
M. The relative standard deviations for 15 repeated measurements of 1
×
10
−6 and 5
×
10
−6
mol
l
−1 HCHO are 4.32 and 3.33%, respectively. The analysis time is 1.5
min. The method was applied to the determination of formaldehyde in urban rainwater. A comparison of results found by proposed method with those obtained by fluorimetric reference method provided a good agreement. |
---|---|
ISSN: | 0039-9140 1873-3573 |
DOI: | 10.1016/j.talanta.2006.05.078 |