Evaluation of the Omron MX3 Plus monitor for blood pressure measurement in adolescents

This study analyzed the accuracy/agreement of the Omron MX3 monitor on 165 adolescents. Blood pressure was measured by the automatic monitor connected in Y with the mercury column (three consecutive and simultaneous measures). The independent measures were analyzed, and the mean differences between...

Full description

Saved in:
Bibliographic Details
Published in:European journal of pediatrics 2009-11, Vol.168 (11), p.1349-1354
Main Authors: Christofaro, Diego Giulliano Destro, Casonatto, Juliano, Polito, Marcos Doederlein, Cardoso, Jefferson Rosa, Fernandes, Rômulo, Guariglia, Débora Alves, Gerage, Aline Mendes, de Oliveira, Arli Ramos
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:This study analyzed the accuracy/agreement of the Omron MX3 monitor on 165 adolescents. Blood pressure was measured by the automatic monitor connected in Y with the mercury column (three consecutive and simultaneous measures). The independent measures were analyzed, and the mean differences between systolic and diastolic measures for both methods were calculated and compared with British Hypertension Society (BHS) and Association for the Advancement of Medical Instrumentation (AAMI) criteria. The automatic monitor received the highest degree of BHS recommendations for systolic and diastolic blood pressures according to the BHS. The median (25th and 75th) difference between the observer and the monitor measurements was −2 (−6 and 1) mmHg for systolic and 0 (−3 and1) mmHg for diastolic pressures. The monitor also satisfies the AAMI standard for the studied population. In conclusion, the Omron MX3 Plus monitor can be considered reliable and valid for clinical practice and is in accordance with BHS and AAMI criteria.