Loading…

The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures

Dose-area product (DAP) measurements were conducted for 168 coronary angiography (CA) and 89 single vessel percutaneous transluminal coronary angioplasty (PTCA) to examine the factors influencing patient dose beyond the X-ray exposure parameters. It was found that for CA, the DAP increases with the...

Full description

Saved in:
Bibliographic Details
Published in:Physica medica 2005-10, Vol.21 (4), p.153-157
Main Authors: Delichas, Miltiadis G., Psarrakos, Kyriakos, Hatziioannou, Konstantinos, Giannoglou, Georgios, Molyvda-Athanasopoulou, Elisabeth, Papanastassiou, Emmanouil, Sioundas, Anastasios
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173
cites cdi_FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173
container_end_page 157
container_issue 4
container_start_page 153
container_title Physica medica
container_volume 21
creator Delichas, Miltiadis G.
Psarrakos, Kyriakos
Hatziioannou, Konstantinos
Giannoglou, Georgios
Molyvda-Athanasopoulou, Elisabeth
Papanastassiou, Emmanouil
Sioundas, Anastasios
description Dose-area product (DAP) measurements were conducted for 168 coronary angiography (CA) and 89 single vessel percutaneous transluminal coronary angioplasty (PTCA) to examine the factors influencing patient dose beyond the X-ray exposure parameters. It was found that for CA, the DAP increases with the number of catheters used and with the number of vessels with stenosis. DAP values for patients with a prior bypass surgery, were higher compared to those without such a medical record to surgery. In PTCA, the use of coronary stents did not enhance the patient radiation dose significantly. Noticeable differences were found in the percentage contribution of each projection to the total DAP between the three types of single vessel PTCA. Finally low variations in DAP were found among the cardiologists performing both procedures.
doi_str_mv 10.1016/S1120-1797(05)80004-3
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_734273465</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S1120179705800043</els_id><sourcerecordid>734273465</sourcerecordid><originalsourceid>FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173</originalsourceid><addsrcrecordid>eNqFkE1vFCEYx0lj09baj6Dhph5GgRlmmJMxG7VNmniwJr0ReHimxczCCGzj9tPLdtf06IHAE_4v8CPkNWcfOOP9xx-cC9bwYRzeMfleMca6pj0iZ3wQXcNHfvuinv9JTsnLnH8x1goh5Qk55artlJLqjDze3CN1uGBwGABpnOhiisdQqIu5zoFOBkpMmSac6024oyXSUl0F4T743xukJjgKcb3M-MeX7S7jKhRMDzXFx2BmujLJ-TjHuy1dUgR0m4T5FTmezJzx4rCfk59fv9ysLpvr79-uVp-vG2j7rjRcKADhLHYTE72Ro0JjAcGqHjtjB2uhH4xAK1FMDgBGBe3Q85Fx64AP7Tl5u8-t1fW1uei1z4DzbALGTdZD24m6elmVcq-EFHNOOOkl-bVJW82Z3lHXT9T1DqlmUj9R1231vTk0bOwa3bPrgLkKPu0FWP_54DHpDH7H2_mEULSL_j8VfwGC65Wj</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>734273465</pqid></control><display><type>article</type><title>The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures</title><source>ScienceDirect Freedom Collection</source><creator>Delichas, Miltiadis G. ; Psarrakos, Kyriakos ; Hatziioannou, Konstantinos ; Giannoglou, Georgios ; Molyvda-Athanasopoulou, Elisabeth ; Papanastassiou, Emmanouil ; Sioundas, Anastasios</creator><creatorcontrib>Delichas, Miltiadis G. ; Psarrakos, Kyriakos ; Hatziioannou, Konstantinos ; Giannoglou, Georgios ; Molyvda-Athanasopoulou, Elisabeth ; Papanastassiou, Emmanouil ; Sioundas, Anastasios</creatorcontrib><description>Dose-area product (DAP) measurements were conducted for 168 coronary angiography (CA) and 89 single vessel percutaneous transluminal coronary angioplasty (PTCA) to examine the factors influencing patient dose beyond the X-ray exposure parameters. It was found that for CA, the DAP increases with the number of catheters used and with the number of vessels with stenosis. DAP values for patients with a prior bypass surgery, were higher compared to those without such a medical record to surgery. In PTCA, the use of coronary stents did not enhance the patient radiation dose significantly. Noticeable differences were found in the percentage contribution of each projection to the total DAP between the three types of single vessel PTCA. Finally low variations in DAP were found among the cardiologists performing both procedures.</description><identifier>ISSN: 1120-1797</identifier><identifier>EISSN: 1724-191X</identifier><identifier>DOI: 10.1016/S1120-1797(05)80004-3</identifier><identifier>PMID: 18348858</identifier><language>eng</language><publisher>Italy: Elsevier Ltd</publisher><subject>Complexity ; Interventional Cardiology ; Radiation dose</subject><ispartof>Physica medica, 2005-10, Vol.21 (4), p.153-157</ispartof><rights>2005 Associazione Italiana di Fisica Medica</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173</citedby><cites>FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/18348858$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Delichas, Miltiadis G.</creatorcontrib><creatorcontrib>Psarrakos, Kyriakos</creatorcontrib><creatorcontrib>Hatziioannou, Konstantinos</creatorcontrib><creatorcontrib>Giannoglou, Georgios</creatorcontrib><creatorcontrib>Molyvda-Athanasopoulou, Elisabeth</creatorcontrib><creatorcontrib>Papanastassiou, Emmanouil</creatorcontrib><creatorcontrib>Sioundas, Anastasios</creatorcontrib><title>The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures</title><title>Physica medica</title><addtitle>Phys Med</addtitle><description>Dose-area product (DAP) measurements were conducted for 168 coronary angiography (CA) and 89 single vessel percutaneous transluminal coronary angioplasty (PTCA) to examine the factors influencing patient dose beyond the X-ray exposure parameters. It was found that for CA, the DAP increases with the number of catheters used and with the number of vessels with stenosis. DAP values for patients with a prior bypass surgery, were higher compared to those without such a medical record to surgery. In PTCA, the use of coronary stents did not enhance the patient radiation dose significantly. Noticeable differences were found in the percentage contribution of each projection to the total DAP between the three types of single vessel PTCA. Finally low variations in DAP were found among the cardiologists performing both procedures.</description><subject>Complexity</subject><subject>Interventional Cardiology</subject><subject>Radiation dose</subject><issn>1120-1797</issn><issn>1724-191X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><recordid>eNqFkE1vFCEYx0lj09baj6Dhph5GgRlmmJMxG7VNmniwJr0ReHimxczCCGzj9tPLdtf06IHAE_4v8CPkNWcfOOP9xx-cC9bwYRzeMfleMca6pj0iZ3wQXcNHfvuinv9JTsnLnH8x1goh5Qk55artlJLqjDze3CN1uGBwGABpnOhiisdQqIu5zoFOBkpMmSac6024oyXSUl0F4T743xukJjgKcb3M-MeX7S7jKhRMDzXFx2BmujLJ-TjHuy1dUgR0m4T5FTmezJzx4rCfk59fv9ysLpvr79-uVp-vG2j7rjRcKADhLHYTE72Ro0JjAcGqHjtjB2uhH4xAK1FMDgBGBe3Q85Fx64AP7Tl5u8-t1fW1uei1z4DzbALGTdZD24m6elmVcq-EFHNOOOkl-bVJW82Z3lHXT9T1DqlmUj9R1231vTk0bOwa3bPrgLkKPu0FWP_54DHpDH7H2_mEULSL_j8VfwGC65Wj</recordid><startdate>20051001</startdate><enddate>20051001</enddate><creator>Delichas, Miltiadis G.</creator><creator>Psarrakos, Kyriakos</creator><creator>Hatziioannou, Konstantinos</creator><creator>Giannoglou, Georgios</creator><creator>Molyvda-Athanasopoulou, Elisabeth</creator><creator>Papanastassiou, Emmanouil</creator><creator>Sioundas, Anastasios</creator><general>Elsevier Ltd</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20051001</creationdate><title>The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures</title><author>Delichas, Miltiadis G. ; Psarrakos, Kyriakos ; Hatziioannou, Konstantinos ; Giannoglou, Georgios ; Molyvda-Athanasopoulou, Elisabeth ; Papanastassiou, Emmanouil ; Sioundas, Anastasios</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>Complexity</topic><topic>Interventional Cardiology</topic><topic>Radiation dose</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Delichas, Miltiadis G.</creatorcontrib><creatorcontrib>Psarrakos, Kyriakos</creatorcontrib><creatorcontrib>Hatziioannou, Konstantinos</creatorcontrib><creatorcontrib>Giannoglou, Georgios</creatorcontrib><creatorcontrib>Molyvda-Athanasopoulou, Elisabeth</creatorcontrib><creatorcontrib>Papanastassiou, Emmanouil</creatorcontrib><creatorcontrib>Sioundas, Anastasios</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Physica medica</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Delichas, Miltiadis G.</au><au>Psarrakos, Kyriakos</au><au>Hatziioannou, Konstantinos</au><au>Giannoglou, Georgios</au><au>Molyvda-Athanasopoulou, Elisabeth</au><au>Papanastassiou, Emmanouil</au><au>Sioundas, Anastasios</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures</atitle><jtitle>Physica medica</jtitle><addtitle>Phys Med</addtitle><date>2005-10-01</date><risdate>2005</risdate><volume>21</volume><issue>4</issue><spage>153</spage><epage>157</epage><pages>153-157</pages><issn>1120-1797</issn><eissn>1724-191X</eissn><abstract>Dose-area product (DAP) measurements were conducted for 168 coronary angiography (CA) and 89 single vessel percutaneous transluminal coronary angioplasty (PTCA) to examine the factors influencing patient dose beyond the X-ray exposure parameters. It was found that for CA, the DAP increases with the number of catheters used and with the number of vessels with stenosis. DAP values for patients with a prior bypass surgery, were higher compared to those without such a medical record to surgery. In PTCA, the use of coronary stents did not enhance the patient radiation dose significantly. Noticeable differences were found in the percentage contribution of each projection to the total DAP between the three types of single vessel PTCA. Finally low variations in DAP were found among the cardiologists performing both procedures.</abstract><cop>Italy</cop><pub>Elsevier Ltd</pub><pmid>18348858</pmid><doi>10.1016/S1120-1797(05)80004-3</doi><tpages>5</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1120-1797
ispartof Physica medica, 2005-10, Vol.21 (4), p.153-157
issn 1120-1797
1724-191X
language eng
recordid cdi_proquest_miscellaneous_734273465
source ScienceDirect Freedom Collection
subjects Complexity
Interventional Cardiology
Radiation dose
title The dependence of patient dose on factors relating to the technique and complexity of Interventional Cardiology procedures
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-06T06%3A21%3A23IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20dependence%20of%20patient%20dose%20on%20factors%20relating%20to%20the%20technique%20and%20complexity%20of%20Interventional%20Cardiology%20procedures&rft.jtitle=Physica%20medica&rft.au=Delichas,%20Miltiadis%20G.&rft.date=2005-10-01&rft.volume=21&rft.issue=4&rft.spage=153&rft.epage=157&rft.pages=153-157&rft.issn=1120-1797&rft.eissn=1724-191X&rft_id=info:doi/10.1016/S1120-1797(05)80004-3&rft_dat=%3Cproquest_cross%3E734273465%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c364t-128cc2dbe4f026a598eabcecb86e4ab7bbc67a2eb5e2fdccc98c3761901bdc173%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=734273465&rft_id=info:pmid/18348858&rfr_iscdi=true