Loading…

One-dimensional photon-counting detector array for use at EUV and soft x-ray wavelengths

The performance of a one-dimensional photon-counting detector array based on microchannel array plates has been evaluated in the laboratory over the wavelength range from 461 A to 1216 A. The output charge pulses from a set of cascaded microchannel array plates were collected on a proximity-focused...

Full description

Saved in:
Bibliographic Details
Published in:Applied optics (2004) 1975-07, Vol.14 (7), p.1632-1644
Main Authors: Timothy, J G, Bybee, R L
Format: Article
Language:English
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The performance of a one-dimensional photon-counting detector array based on microchannel array plates has been evaluated in the laboratory over the wavelength range from 461 A to 1216 A. The output charge pulses from a set of cascaded microchannel array plates were collected on a proximity-focused anode array consisting of sixty-four linear electrodes 1.3 mm in length and 25 microm in width, set on 50-microm centers. Although the dynamic range of the detector array was severely limited by ion-feedback effects in currently available microchannel array plates, the theoretical spatial resolution of 50 microm was achieved with a detection efficiency at wavelengths below 800 A equal to that of a conventional channel electron multiplier.
ISSN:1559-128X
DOI:10.1364/AO.14.001632