Loading…
One-dimensional photon-counting detector array for use at EUV and soft x-ray wavelengths
The performance of a one-dimensional photon-counting detector array based on microchannel array plates has been evaluated in the laboratory over the wavelength range from 461 A to 1216 A. The output charge pulses from a set of cascaded microchannel array plates were collected on a proximity-focused...
Saved in:
Published in: | Applied optics (2004) 1975-07, Vol.14 (7), p.1632-1644 |
---|---|
Main Authors: | , |
Format: | Article |
Language: | English |
Citations: | Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The performance of a one-dimensional photon-counting detector array based on microchannel array plates has been evaluated in the laboratory over the wavelength range from 461 A to 1216 A. The output charge pulses from a set of cascaded microchannel array plates were collected on a proximity-focused anode array consisting of sixty-four linear electrodes 1.3 mm in length and 25 microm in width, set on 50-microm centers. Although the dynamic range of the detector array was severely limited by ion-feedback effects in currently available microchannel array plates, the theoretical spatial resolution of 50 microm was achieved with a detection efficiency at wavelengths below 800 A equal to that of a conventional channel electron multiplier. |
---|---|
ISSN: | 1559-128X |
DOI: | 10.1364/AO.14.001632 |