Effects of acetabular abduction on cup wear rates in total hip arthroplasty

This study was designed to determine the effect of acetabular abduction on the polyethylene wear rates of the acetabular component. The hypothesis of this study is that acetabular placement, in particular abduction, effects contact forces and therefore polyethylene wear. A total of 364 total hip art...

Full description

Saved in:
Bibliographic Details
Published in:The Journal of arthroplasty 1998-09, Vol.13 (6), p.621-626
Main Authors: Del Schutte, H., Lipman, Andrew J., Bannar, Stephen M., Livermore, John T., Ilstrup, Duane, Morrey, Bernard F.
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63
cites cdi_FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63
container_end_page 626
container_issue 6
container_start_page 621
container_title The Journal of arthroplasty
container_volume 13
creator Del Schutte, H.
Lipman, Andrew J.
Bannar, Stephen M.
Livermore, John T.
Ilstrup, Duane
Morrey, Bernard F.
description This study was designed to determine the effect of acetabular abduction on the polyethylene wear rates of the acetabular component. The hypothesis of this study is that acetabular placement, in particular abduction, effects contact forces and therefore polyethylene wear. A total of 364 total hip arthroplasties that were performed between 1974 and 1976 were included in this study. The cemented all-polyethylene acetabular components were the same for each case and came from a single supplier. The polyethylene wear rates were calculated by measuring the shortest radius from the center of the prosthetic femoral head to a point on the outer surface of the acetabular cup. An immediate postoperative radiograph was compared with a follow-up radiograph at least 9.5 years later. A standardized radiograph was used to prevent differences due to magnification. The abduction or inclination of the acetabular cups was measured in all of the patients. The amount of acetabular cup abduction was measured relative to the ischial tuberosity line. The mean abduction was 44.1° with a standard deviation of 9.2° and a median of 44°. The range of abduction was from 0° to 85°. This analysis failed to show a correlation between the amount of acetabular abduction and polyethylene wear rates (Pearson's correlation coefficient = 0.0679; P = .20). With a sample size of 364, there is over a 95% change (statistical power) of detecting an underlying true correlation between wear and abduction that is greater than or equal to 0.20. We were unable to demonstrate a difference in wear that would correlate with the differing degrees of abduction of the acetabular socket. We felt that the radiographic measurements of wear were quite accurate. This method of determining acetabular abduction has also been documented and supported in previous literature and has allowed us to accurately determine acetabular abduction. The results of our study demonstrate that within the normal ranges described by our study, polyethylene wear did not significantly increase with increased abduction of the acetabular component.
doi_str_mv 10.1016/S0883-5403(98)80003-X
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_73915501</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S088354039880003X</els_id><sourcerecordid>73915501</sourcerecordid><originalsourceid>FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63</originalsourceid><addsrcrecordid>eNqFkMtKBDEQRYMoOj4-QchKdNFamTw6vRIZfKHgQgV3IZNUMNIz3SZpxb-3dQa3QkFB1b11qUPIIYNTBkydPYLWvJIC-HGjTzQA8Oplg0yY5NNKC1CbZPIn2SG7Ob8BMCal2CbbTS2Y4GpC7i5DQFcy7QK1DoudD61N1M794ErslnQsN_T0E8dpsgUzjUtaumJb-hp7alN5TV3f2ly-9slWsG3Gg3XfI89Xl0-zm-r-4fp2dnFfOTFVpZor59EpoTmA500DtUDppz64AEE5LWplhZLOaVZL6bHmFiV3DSgMWAfF98jR6m6fuvcBczGLmB22rV1iN2RT82Z8E9golCuhS13OCYPpU1zY9GUYmB-I5hei-SFkGm1-IZqX0Xe4DhjmC_R_rjW1cX--2uP45UfEZLKLuHToYxphGt_FfxK-AVp_ggs</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>73915501</pqid></control><display><type>article</type><title>Effects of acetabular abduction on cup wear rates in total hip arthroplasty</title><source>Elsevier</source><creator>Del Schutte, H. ; Lipman, Andrew J. ; Bannar, Stephen M. ; Livermore, John T. ; Ilstrup, Duane ; Morrey, Bernard F.</creator><creatorcontrib>Del Schutte, H. ; Lipman, Andrew J. ; Bannar, Stephen M. ; Livermore, John T. ; Ilstrup, Duane ; Morrey, Bernard F.</creatorcontrib><description>This study was designed to determine the effect of acetabular abduction on the polyethylene wear rates of the acetabular component. The hypothesis of this study is that acetabular placement, in particular abduction, effects contact forces and therefore polyethylene wear. A total of 364 total hip arthroplasties that were performed between 1974 and 1976 were included in this study. The cemented all-polyethylene acetabular components were the same for each case and came from a single supplier. The polyethylene wear rates were calculated by measuring the shortest radius from the center of the prosthetic femoral head to a point on the outer surface of the acetabular cup. An immediate postoperative radiograph was compared with a follow-up radiograph at least 9.5 years later. A standardized radiograph was used to prevent differences due to magnification. The abduction or inclination of the acetabular cups was measured in all of the patients. The amount of acetabular cup abduction was measured relative to the ischial tuberosity line. The mean abduction was 44.1° with a standard deviation of 9.2° and a median of 44°. The range of abduction was from 0° to 85°. This analysis failed to show a correlation between the amount of acetabular abduction and polyethylene wear rates (Pearson's correlation coefficient = 0.0679; P = .20). With a sample size of 364, there is over a 95% change (statistical power) of detecting an underlying true correlation between wear and abduction that is greater than or equal to 0.20. We were unable to demonstrate a difference in wear that would correlate with the differing degrees of abduction of the acetabular socket. We felt that the radiographic measurements of wear were quite accurate. This method of determining acetabular abduction has also been documented and supported in previous literature and has allowed us to accurately determine acetabular abduction. The results of our study demonstrate that within the normal ranges described by our study, polyethylene wear did not significantly increase with increased abduction of the acetabular component.</description><identifier>ISSN: 0883-5403</identifier><identifier>EISSN: 1532-8406</identifier><identifier>DOI: 10.1016/S0883-5403(98)80003-X</identifier><identifier>PMID: 9741436</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>acetabular abduction ; Acetabulum - diagnostic imaging ; Acetabulum - surgery ; Biomechanical Phenomena ; Equipment Failure Analysis ; Follow-Up Studies ; hip arthroplasty ; Hip Prosthesis ; Humans ; polyethylene wear rates ; Polyethylenes ; Postoperative Complications - diagnostic imaging ; Prosthesis Design ; Radiography ; Range of Motion, Articular - physiology ; Surface Properties</subject><ispartof>The Journal of arthroplasty, 1998-09, Vol.13 (6), p.621-626</ispartof><rights>1998 Churchill Livingstone. All rights reserved</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63</citedby><cites>FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/9741436$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Del Schutte, H.</creatorcontrib><creatorcontrib>Lipman, Andrew J.</creatorcontrib><creatorcontrib>Bannar, Stephen M.</creatorcontrib><creatorcontrib>Livermore, John T.</creatorcontrib><creatorcontrib>Ilstrup, Duane</creatorcontrib><creatorcontrib>Morrey, Bernard F.</creatorcontrib><title>Effects of acetabular abduction on cup wear rates in total hip arthroplasty</title><title>The Journal of arthroplasty</title><addtitle>J Arthroplasty</addtitle><description>This study was designed to determine the effect of acetabular abduction on the polyethylene wear rates of the acetabular component. The hypothesis of this study is that acetabular placement, in particular abduction, effects contact forces and therefore polyethylene wear. A total of 364 total hip arthroplasties that were performed between 1974 and 1976 were included in this study. The cemented all-polyethylene acetabular components were the same for each case and came from a single supplier. The polyethylene wear rates were calculated by measuring the shortest radius from the center of the prosthetic femoral head to a point on the outer surface of the acetabular cup. An immediate postoperative radiograph was compared with a follow-up radiograph at least 9.5 years later. A standardized radiograph was used to prevent differences due to magnification. The abduction or inclination of the acetabular cups was measured in all of the patients. The amount of acetabular cup abduction was measured relative to the ischial tuberosity line. The mean abduction was 44.1° with a standard deviation of 9.2° and a median of 44°. The range of abduction was from 0° to 85°. This analysis failed to show a correlation between the amount of acetabular abduction and polyethylene wear rates (Pearson's correlation coefficient = 0.0679; P = .20). With a sample size of 364, there is over a 95% change (statistical power) of detecting an underlying true correlation between wear and abduction that is greater than or equal to 0.20. We were unable to demonstrate a difference in wear that would correlate with the differing degrees of abduction of the acetabular socket. We felt that the radiographic measurements of wear were quite accurate. This method of determining acetabular abduction has also been documented and supported in previous literature and has allowed us to accurately determine acetabular abduction. The results of our study demonstrate that within the normal ranges described by our study, polyethylene wear did not significantly increase with increased abduction of the acetabular component.</description><subject>acetabular abduction</subject><subject>Acetabulum - diagnostic imaging</subject><subject>Acetabulum - surgery</subject><subject>Biomechanical Phenomena</subject><subject>Equipment Failure Analysis</subject><subject>Follow-Up Studies</subject><subject>hip arthroplasty</subject><subject>Hip Prosthesis</subject><subject>Humans</subject><subject>polyethylene wear rates</subject><subject>Polyethylenes</subject><subject>Postoperative Complications - diagnostic imaging</subject><subject>Prosthesis Design</subject><subject>Radiography</subject><subject>Range of Motion, Articular - physiology</subject><subject>Surface Properties</subject><issn>0883-5403</issn><issn>1532-8406</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1998</creationdate><recordtype>article</recordtype><recordid>eNqFkMtKBDEQRYMoOj4-QchKdNFamTw6vRIZfKHgQgV3IZNUMNIz3SZpxb-3dQa3QkFB1b11qUPIIYNTBkydPYLWvJIC-HGjTzQA8Oplg0yY5NNKC1CbZPIn2SG7Ob8BMCal2CbbTS2Y4GpC7i5DQFcy7QK1DoudD61N1M794ErslnQsN_T0E8dpsgUzjUtaumJb-hp7alN5TV3f2ly-9slWsG3Gg3XfI89Xl0-zm-r-4fp2dnFfOTFVpZor59EpoTmA500DtUDppz64AEE5LWplhZLOaVZL6bHmFiV3DSgMWAfF98jR6m6fuvcBczGLmB22rV1iN2RT82Z8E9golCuhS13OCYPpU1zY9GUYmB-I5hei-SFkGm1-IZqX0Xe4DhjmC_R_rjW1cX--2uP45UfEZLKLuHToYxphGt_FfxK-AVp_ggs</recordid><startdate>19980901</startdate><enddate>19980901</enddate><creator>Del Schutte, H.</creator><creator>Lipman, Andrew J.</creator><creator>Bannar, Stephen M.</creator><creator>Livermore, John T.</creator><creator>Ilstrup, Duane</creator><creator>Morrey, Bernard F.</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19980901</creationdate><title>Effects of acetabular abduction on cup wear rates in total hip arthroplasty</title><author>Del Schutte, H. ; Lipman, Andrew J. ; Bannar, Stephen M. ; Livermore, John T. ; Ilstrup, Duane ; Morrey, Bernard F.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1998</creationdate><topic>acetabular abduction</topic><topic>Acetabulum - diagnostic imaging</topic><topic>Acetabulum - surgery</topic><topic>Biomechanical Phenomena</topic><topic>Equipment Failure Analysis</topic><topic>Follow-Up Studies</topic><topic>hip arthroplasty</topic><topic>Hip Prosthesis</topic><topic>Humans</topic><topic>polyethylene wear rates</topic><topic>Polyethylenes</topic><topic>Postoperative Complications - diagnostic imaging</topic><topic>Prosthesis Design</topic><topic>Radiography</topic><topic>Range of Motion, Articular - physiology</topic><topic>Surface Properties</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Del Schutte, H.</creatorcontrib><creatorcontrib>Lipman, Andrew J.</creatorcontrib><creatorcontrib>Bannar, Stephen M.</creatorcontrib><creatorcontrib>Livermore, John T.</creatorcontrib><creatorcontrib>Ilstrup, Duane</creatorcontrib><creatorcontrib>Morrey, Bernard F.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of arthroplasty</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Del Schutte, H.</au><au>Lipman, Andrew J.</au><au>Bannar, Stephen M.</au><au>Livermore, John T.</au><au>Ilstrup, Duane</au><au>Morrey, Bernard F.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Effects of acetabular abduction on cup wear rates in total hip arthroplasty</atitle><jtitle>The Journal of arthroplasty</jtitle><addtitle>J Arthroplasty</addtitle><date>1998-09-01</date><risdate>1998</risdate><volume>13</volume><issue>6</issue><spage>621</spage><epage>626</epage><pages>621-626</pages><issn>0883-5403</issn><eissn>1532-8406</eissn><abstract>This study was designed to determine the effect of acetabular abduction on the polyethylene wear rates of the acetabular component. The hypothesis of this study is that acetabular placement, in particular abduction, effects contact forces and therefore polyethylene wear. A total of 364 total hip arthroplasties that were performed between 1974 and 1976 were included in this study. The cemented all-polyethylene acetabular components were the same for each case and came from a single supplier. The polyethylene wear rates were calculated by measuring the shortest radius from the center of the prosthetic femoral head to a point on the outer surface of the acetabular cup. An immediate postoperative radiograph was compared with a follow-up radiograph at least 9.5 years later. A standardized radiograph was used to prevent differences due to magnification. The abduction or inclination of the acetabular cups was measured in all of the patients. The amount of acetabular cup abduction was measured relative to the ischial tuberosity line. The mean abduction was 44.1° with a standard deviation of 9.2° and a median of 44°. The range of abduction was from 0° to 85°. This analysis failed to show a correlation between the amount of acetabular abduction and polyethylene wear rates (Pearson's correlation coefficient = 0.0679; P = .20). With a sample size of 364, there is over a 95% change (statistical power) of detecting an underlying true correlation between wear and abduction that is greater than or equal to 0.20. We were unable to demonstrate a difference in wear that would correlate with the differing degrees of abduction of the acetabular socket. We felt that the radiographic measurements of wear were quite accurate. This method of determining acetabular abduction has also been documented and supported in previous literature and has allowed us to accurately determine acetabular abduction. The results of our study demonstrate that within the normal ranges described by our study, polyethylene wear did not significantly increase with increased abduction of the acetabular component.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>9741436</pmid><doi>10.1016/S0883-5403(98)80003-X</doi><tpages>6</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0883-5403
ispartof The Journal of arthroplasty, 1998-09, Vol.13 (6), p.621-626
issn 0883-5403
1532-8406
language eng
recordid cdi_proquest_miscellaneous_73915501
source Elsevier
subjects acetabular abduction
Acetabulum - diagnostic imaging
Acetabulum - surgery
Biomechanical Phenomena
Equipment Failure Analysis
Follow-Up Studies
hip arthroplasty
Hip Prosthesis
Humans
polyethylene wear rates
Polyethylenes
Postoperative Complications - diagnostic imaging
Prosthesis Design
Radiography
Range of Motion, Articular - physiology
Surface Properties
title Effects of acetabular abduction on cup wear rates in total hip arthroplasty
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-07T20%3A33%3A43IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Effects%20of%20acetabular%20abduction%20on%20cup%20wear%20rates%20in%20total%20hip%20arthroplasty&rft.jtitle=The%20Journal%20of%20arthroplasty&rft.au=Del%20Schutte,%20H.&rft.date=1998-09-01&rft.volume=13&rft.issue=6&rft.spage=621&rft.epage=626&rft.pages=621-626&rft.issn=0883-5403&rft.eissn=1532-8406&rft_id=info:doi/10.1016/S0883-5403(98)80003-X&rft_dat=%3Cproquest_cross%3E73915501%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c426t-b6cdec648300d399074e5d2dfcf0f6c8476a465cc81755de73ae53c906efe7f63%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=73915501&rft_id=info:pmid/9741436&rfr_iscdi=true