Loading…

Comparison of a prototype esophageal oximetry probe with two conventional digital pulse oximetry monitors in aortocoronary bypass patients

Pulse oximetry (SpO2) is the non-invasive standard for monitoring arterial oxygen saturation in patients undergoing anesthesia, but is subject to external interference by motion artifact, peripheral vasoconstriction, and low cardiac output. We hypothesized that oximetry signals could be acquired fro...

Full description

Saved in:
Bibliographic Details
Published in:Journal of clinical monitoring and computing 2000, Vol.16 (3), p.201-209
Main Authors: PRICLIPP, Richard C, SCUDERI, Phillip E, HINES, Michael H, ATLEE, John L, BUTTERWORTH, John F
Format: Article
Language:English
Subjects:
Citations: Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Pulse oximetry (SpO2) is the non-invasive standard for monitoring arterial oxygen saturation in patients undergoing anesthesia, but is subject to external interference by motion artifact, peripheral vasoconstriction, and low cardiac output. We hypothesized that oximetry signals could be acquired from the esophagus when peripheral pulse oximetry is unobtainable. Therefore, we tested an esophageal stethoscope which incorporates transverse oximetry photodetectors and emitters in patients undergoing coronary bypass surgery. Immediately after induction of general anesthesia in 10 coronary artery bypass (CABG) patients, Criticare and Nellcor digital probes were positioned on the left hand, concurrent with placement of an esophageal SpO2 probe. A computer recorded 5,910 matched oximetry signals every 15 sec during an average of 2.5 hrs. All SpO2 measurements were before, and immediately after non-pulsatile, hypothermic cardiopulmonary bypass. Data represent the percentage (median value [range]) of the total monitored time that a SpO2 value was displayed. The Nellcor (99.8%, range 6.5-100%) and Criticare (99.7%, range 36.6-100%) acquired and displayed saturation signals more frequently (p = 0.003) than the esophageal monitor (75.3%, range 42.1-95.8%). The two standard digital oximeters had a mean difference of 0.9%, with a standard deviation of the differences of 0.9. The esophageal probe had a mean difference of -5.2% and -4.8%, with standard deviation of differences of 8.0 and 7.7 (compared to the Nellcor and Criticare monitors, respectively). A second-generation prototype shielded from electrocautery interference was tested in an additional 4 patients. The shielded prototype displayed signals more frequently (96.7%, range 68.4-100%) than the original esophageal prototype. Digital pulse oximetry failure is common in CABG patients, probably because of marginal cardiac output and peripheral vasoconstriction associated with hypothermia. Our study could not confirm that esophageal technology, which utilizes the esophagus as a site of transflectance oximetry, was superior to conventional digital pulse oximetry.
ISSN:1387-1307
1573-2614
DOI:10.1023/A:1009941610320