Loading…
Evaluation of the primary frequency standard NPL-CsF1
A new caesium fountain frequency standard (NPL-CsF1) at the National Physical Laboratory is described. Procedures for evaluation of the systematic frequency shifts are presented. The NPL-CsF1 has a short-term stability sy(t) = 1.4 X 10-13t-1/2, predominantly due to the local oscillator phase noise....
Saved in:
Published in: | Metrologia 2005-02, Vol.42 (1), p.49-57 |
---|---|
Main Authors: | , , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | A new caesium fountain frequency standard (NPL-CsF1) at the National Physical Laboratory is described. Procedures for evaluation of the systematic frequency shifts are presented. The NPL-CsF1 has a short-term stability sy(t) = 1.4 X 10-13t-1/2, predominantly due to the local oscillator phase noise. The accuracy of 1 part in 1015 is limited by the uncertainty of the frequency shift due to collisions between cold atoms. |
---|---|
ISSN: | 0026-1394 1681-7575 |
DOI: | 10.1088/0026-1394/42/1/007 |