Physical and biological characterization of a seawater ultraviolet radiation sterilizer
The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array o...
Saved in:
Published in: | Radiation physics and chemistry (Oxford, England : 1993) England : 1993), 1994, Vol.43 (3), p.249-255 |
---|---|
Main Authors: | , , , |
Format: | Article |
Language: | English |
Citations: | Items that this one cites Items that cite this one |
Online Access: | Get full text |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array of adjacent lamps, and by measuring the UV radiation absorption of seawater with respect to distilled water. The biological characterization was performed by measuring the amount of reduction of bacteria in stored seawater after different filtration and UV treatments. Among the filtration methods tested, differential filtration (5, 3 and 0.45 μm filters connected in series) caused the highest bacterial reduction factor of 60%. UV radiant exposures of 212, 424, 636 and 848 J m
−2 yielded bacteria reduction factors of 99.86, 99.969, 99.997 and 100%, respectively, for populations of
Vibrio and
Pseudomonas bacteria present in stored seawater. It is concluded that the system is useful for water disinfection when 1, 2 or 3 lamps are on; when 4 lamps are used the treated water becomes sterile. |
---|---|