Physical and biological characterization of a seawater ultraviolet radiation sterilizer

The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array o...

Full description

Saved in:
Bibliographic Details
Published in:Radiation physics and chemistry (Oxford, England : 1993) England : 1993), 1994, Vol.43 (3), p.249-255
Main Authors: Torrentera, Laura, Uribe, Roberto M., Rodríguez, Romana R., Carrillo, Ricardo E.
Format: Article
Language:English
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
cited_by cdi_FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3
cites cdi_FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3
container_end_page 255
container_issue 3
container_start_page 249
container_title Radiation physics and chemistry (Oxford, England : 1993)
container_volume 43
creator Torrentera, Laura
Uribe, Roberto M.
Rodríguez, Romana R.
Carrillo, Ricardo E.
description The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array of adjacent lamps, and by measuring the UV radiation absorption of seawater with respect to distilled water. The biological characterization was performed by measuring the amount of reduction of bacteria in stored seawater after different filtration and UV treatments. Among the filtration methods tested, differential filtration (5, 3 and 0.45 μm filters connected in series) caused the highest bacterial reduction factor of 60%. UV radiant exposures of 212, 424, 636 and 848 J m −2 yielded bacteria reduction factors of 99.86, 99.969, 99.997 and 100%, respectively, for populations of Vibrio and Pseudomonas bacteria present in stored seawater. It is concluded that the system is useful for water disinfection when 1, 2 or 3 lamps are on; when 4 lamps are used the treated water becomes sterile.
doi_str_mv 10.1016/0969-806X(94)90187-2
format article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_745653249</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>0969806X94901872</els_id><sourcerecordid>745653249</sourcerecordid><originalsourceid>FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3</originalsourceid><addsrcrecordid>eNp9kD1PwzAQhi0EEqXwDxiyAUPAThzHXpBQxZdUCQYQbNbZuVCjNC52WtT-epIWMTKd7vS8r3QPIaeMXjLKxBVVQqWSivdzxS8UZbJMsz0y6qdKqVTFPhn9IYfkKMZPSmkpi3xE3p5n6-gsNAm0VWKcb_zHdrUzCGA7DG4DnfNt4usEkojwDf0xWTZdgFWPY5cEqNyOiQPfuA2GY3JQQxPx5HeOyevd7cvkIZ0-3T9ObqapzXPRpZKhpBIEQ-CK57U0EowSAKyuqClAGVYZzpRgJhOVMLa2xmamVJybgpc2H5OzXe8i-K8lxk7PXbTYNNCiX0Zd8kIUecZVT_IdaYOPMWCtF8HNIaw1o3rQqAdHenCkFddbjTrrY9e7GPZfrBwGHa3D1mLlAtpOV979X_AD5J18fA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>745653249</pqid></control><display><type>article</type><title>Physical and biological characterization of a seawater ultraviolet radiation sterilizer</title><source>Backfile Package - Physics General (Legacy) [YPA]</source><creator>Torrentera, Laura ; Uribe, Roberto M. ; Rodríguez, Romana R. ; Carrillo, Ricardo E.</creator><creatorcontrib>Torrentera, Laura ; Uribe, Roberto M. ; Rodríguez, Romana R. ; Carrillo, Ricardo E.</creatorcontrib><description>The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array of adjacent lamps, and by measuring the UV radiation absorption of seawater with respect to distilled water. The biological characterization was performed by measuring the amount of reduction of bacteria in stored seawater after different filtration and UV treatments. Among the filtration methods tested, differential filtration (5, 3 and 0.45 μm filters connected in series) caused the highest bacterial reduction factor of 60%. UV radiant exposures of 212, 424, 636 and 848 J m −2 yielded bacteria reduction factors of 99.86, 99.969, 99.997 and 100%, respectively, for populations of Vibrio and Pseudomonas bacteria present in stored seawater. It is concluded that the system is useful for water disinfection when 1, 2 or 3 lamps are on; when 4 lamps are used the treated water becomes sterile.</description><identifier>ISSN: 0969-806X</identifier><identifier>EISSN: 1879-0895</identifier><identifier>DOI: 10.1016/0969-806X(94)90187-2</identifier><language>eng</language><publisher>Elsevier Ltd</publisher><ispartof>Radiation physics and chemistry (Oxford, England : 1993), 1994, Vol.43 (3), p.249-255</ispartof><rights>1994</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3</citedby><cites>FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.sciencedirect.com/science/article/pii/0969806X94901872$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,776,780,3618,4009,27902,27903,27904,45990</link.rule.ids></links><search><creatorcontrib>Torrentera, Laura</creatorcontrib><creatorcontrib>Uribe, Roberto M.</creatorcontrib><creatorcontrib>Rodríguez, Romana R.</creatorcontrib><creatorcontrib>Carrillo, Ricardo E.</creatorcontrib><title>Physical and biological characterization of a seawater ultraviolet radiation sterilizer</title><title>Radiation physics and chemistry (Oxford, England : 1993)</title><description>The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array of adjacent lamps, and by measuring the UV radiation absorption of seawater with respect to distilled water. The biological characterization was performed by measuring the amount of reduction of bacteria in stored seawater after different filtration and UV treatments. Among the filtration methods tested, differential filtration (5, 3 and 0.45 μm filters connected in series) caused the highest bacterial reduction factor of 60%. UV radiant exposures of 212, 424, 636 and 848 J m −2 yielded bacteria reduction factors of 99.86, 99.969, 99.997 and 100%, respectively, for populations of Vibrio and Pseudomonas bacteria present in stored seawater. It is concluded that the system is useful for water disinfection when 1, 2 or 3 lamps are on; when 4 lamps are used the treated water becomes sterile.</description><issn>0969-806X</issn><issn>1879-0895</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1994</creationdate><recordtype>article</recordtype><recordid>eNp9kD1PwzAQhi0EEqXwDxiyAUPAThzHXpBQxZdUCQYQbNbZuVCjNC52WtT-epIWMTKd7vS8r3QPIaeMXjLKxBVVQqWSivdzxS8UZbJMsz0y6qdKqVTFPhn9IYfkKMZPSmkpi3xE3p5n6-gsNAm0VWKcb_zHdrUzCGA7DG4DnfNt4usEkojwDf0xWTZdgFWPY5cEqNyOiQPfuA2GY3JQQxPx5HeOyevd7cvkIZ0-3T9ObqapzXPRpZKhpBIEQ-CK57U0EowSAKyuqClAGVYZzpRgJhOVMLa2xmamVJybgpc2H5OzXe8i-K8lxk7PXbTYNNCiX0Zd8kIUecZVT_IdaYOPMWCtF8HNIaw1o3rQqAdHenCkFddbjTrrY9e7GPZfrBwGHa3D1mLlAtpOV979X_AD5J18fA</recordid><startdate>1994</startdate><enddate>1994</enddate><creator>Torrentera, Laura</creator><creator>Uribe, Roberto M.</creator><creator>Rodríguez, Romana R.</creator><creator>Carrillo, Ricardo E.</creator><general>Elsevier Ltd</general><scope>AAYXX</scope><scope>CITATION</scope><scope>7QO</scope><scope>8FD</scope><scope>FR3</scope><scope>P64</scope></search><sort><creationdate>1994</creationdate><title>Physical and biological characterization of a seawater ultraviolet radiation sterilizer</title><author>Torrentera, Laura ; Uribe, Roberto M. ; Rodríguez, Romana R. ; Carrillo, Ricardo E.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1994</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Torrentera, Laura</creatorcontrib><creatorcontrib>Uribe, Roberto M.</creatorcontrib><creatorcontrib>Rodríguez, Romana R.</creatorcontrib><creatorcontrib>Carrillo, Ricardo E.</creatorcontrib><collection>CrossRef</collection><collection>Biotechnology Research Abstracts</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><jtitle>Radiation physics and chemistry (Oxford, England : 1993)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Torrentera, Laura</au><au>Uribe, Roberto M.</au><au>Rodríguez, Romana R.</au><au>Carrillo, Ricardo E.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Physical and biological characterization of a seawater ultraviolet radiation sterilizer</atitle><jtitle>Radiation physics and chemistry (Oxford, England : 1993)</jtitle><date>1994</date><risdate>1994</risdate><volume>43</volume><issue>3</issue><spage>249</spage><epage>255</epage><pages>249-255</pages><issn>0969-806X</issn><eissn>1879-0895</eissn><abstract>The physical and biological characterization of a seawater ultraviolet (UV) sterilizer is described. The physical characterization was performed using radiochromic dye films by evaluating the uniformity of the radiant exposure along each lamp, the effect of the radiation from one lamp on the array of adjacent lamps, and by measuring the UV radiation absorption of seawater with respect to distilled water. The biological characterization was performed by measuring the amount of reduction of bacteria in stored seawater after different filtration and UV treatments. Among the filtration methods tested, differential filtration (5, 3 and 0.45 μm filters connected in series) caused the highest bacterial reduction factor of 60%. UV radiant exposures of 212, 424, 636 and 848 J m −2 yielded bacteria reduction factors of 99.86, 99.969, 99.997 and 100%, respectively, for populations of Vibrio and Pseudomonas bacteria present in stored seawater. It is concluded that the system is useful for water disinfection when 1, 2 or 3 lamps are on; when 4 lamps are used the treated water becomes sterile.</abstract><pub>Elsevier Ltd</pub><doi>10.1016/0969-806X(94)90187-2</doi><tpages>7</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0969-806X
ispartof Radiation physics and chemistry (Oxford, England : 1993), 1994, Vol.43 (3), p.249-255
issn 0969-806X
1879-0895
language eng
recordid cdi_proquest_miscellaneous_745653249
source Backfile Package - Physics General (Legacy) [YPA]
title Physical and biological characterization of a seawater ultraviolet radiation sterilizer
url http://sfxeu10.hosted.exlibrisgroup.com/loughborough?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-25T21%3A50%3A07IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Physical%20and%20biological%20characterization%20of%20a%20seawater%20ultraviolet%20radiation%20sterilizer&rft.jtitle=Radiation%20physics%20and%20chemistry%20(Oxford,%20England%20:%201993)&rft.au=Torrentera,%20Laura&rft.date=1994&rft.volume=43&rft.issue=3&rft.spage=249&rft.epage=255&rft.pages=249-255&rft.issn=0969-806X&rft.eissn=1879-0895&rft_id=info:doi/10.1016/0969-806X(94)90187-2&rft_dat=%3Cproquest_cross%3E745653249%3C/proquest_cross%3E%3Cgrp_id%3Ecdi_FETCH-LOGICAL-c336t-81e808a61ea4943f8b8ab96aa1fd0b5a9b1db41961b26d6bcfcbc2b7944b547c3%3C/grp_id%3E%3Coa%3E%3C/oa%3E%3Curl%3E%3C/url%3E&rft_id=info:oai/&rft_pqid=745653249&rft_id=info:pmid/&rfr_iscdi=true