The patellar tendon bar! Is it a necessary feature?

Abstract The purpose of this investigation was to vary the load on the patellar tendon bar and to study the subsequent effect this has on the pattern of the pressure distribution at the stump–socket interface. Ten male subjects from the Southern General Hospital in Glasgow, UK participated in this s...

Full description

Saved in:
Bibliographic Details
Published in:Medical engineering & physics 2010-09, Vol.32 (7), p.760-765
Main Authors: Abu Osman, N.A, Spence, W.D, Solomonidis, S.E, Paul, J.P, Weir, A.M
Format: Article
Language:English
Subjects:
Citations: Items that this one cites
Items that cite this one
Online Access:Get full text
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Abstract The purpose of this investigation was to vary the load on the patellar tendon bar and to study the subsequent effect this has on the pattern of the pressure distribution at the stump–socket interface. Ten male subjects from the Southern General Hospital in Glasgow, UK participated in this study. Measuring systems utilising strain gauge and electrohydraulic technologies were designed, developed and constructed to enable pressure measurements to be conducted. One transducer, the patellar tendon (PT) transducer, was attached to the patellar tendon bar of the socket such that the patellar tendon bar was capable of being translated by ±10 mm towards or away from the tendon. The results of this study showed that the position of the patellar tendon bar had no significant effect on the pressure distribution around the socket indicating that it is an unnecessary feature, which, we propose, may be eliminated during manufacture of a trans-tibial socket.
ISSN:1350-4533
1873-4030
DOI:10.1016/j.medengphy.2010.04.020